Objectives: Understanding the severity of aortic regurgitation (AR) after transcatheter aortic valve implantation, its impact on left ventricular (LV) structure and function, and the structural factors associated with worsening AR could lead to improvements in patient selection, implantation technique, and valve design.

Background: Initial studies in patients at high risk of surgical aortic valve replacement have reported both central valvular and paravalvular AR after transcatheter aortic valve implantation.

Methods: Transthoracic echocardiograms were quantified from 95 patients in the REVIVAL (TRanscatheter EndoVascular Implantation of VALves) trial. Transthoracic echocardiograms were obtained before implantation of the Edwards-Sapien valve (Edwards Lifesciences, Irvine, California) and thereafter at selected intervals. Measurements included LV internal diameters and volumes, ejection fraction, aortic valve area, and the degree of aortic regurgitation. Measures of degree of native leaflet mobility, thickness, and calcification, as well as left ventricular outflow tract, aortic annulus, and aortic root diameters were also made.

Results: Eighty-four patients remained after 11 were excluded; 26 (29.8%) died over a period of 3 years. At 24 h post-implantation, 75% had some degree of AR, mostly paravalvular. By 1 year, the mean AR grade increased slightly, but not significantly (1.1 ± 0.8 to 1.3 ± 0.9), and all measures of LV structure and function improved (LV ejection fraction, 50.7 ± 16.1% to 59.4 ± 14.0%). Native aortic leaflet calcification and annulus diameter correlated significantly with the severity of AR at 1 year (p < 0.05).

Conclusions: AR after transcatheter aortic valve implantation is frequent but is rarely more than mild. Although AR progresses, it is not associated with a harmful impact on LV structure and function over the first year. Native valve calcification and aortic annulus diameter influence the degree of AR at 6 months.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcmg.2012.02.008DOI Listing

Publication Analysis

Top Keywords

aortic valve
24
transcatheter aortic
16
aortic regurgitation
12
aortic
12
valve implantation
12
structure function
12
valve
9
regurgitation transcatheter
8
left ventricular
8
transthoracic echocardiograms
8

Similar Publications

Background: Evidence regarding the incidence of prosthesis-patient mismatch (PPM) and long-term mortality after transcatheter aortic valve replacement (TAVR) in patients with bicuspid aortic valve stenosis (AS) is scarce.

Objectives: This study sought to assess the incidence and prognostic impact of PPM after TAVR for bicuspid AS compared with that for tricuspid AS.

Methods: In total, 7,393 patients who underwent TAVR were prospectively enrolled in the OCEAN-TAVI (Optimized Catheter Valvular Intervention Transcatheter Aortic Valve Implantation) registry, an ongoing Japanese, multicenter registry.

View Article and Find Full Text PDF

Revisiting the Neuropsychological and Clinical Profile of Mosaic Turner Syndrome With a Ring X Chromosome.

Am J Med Genet C Semin Med Genet

January 2025

Medical Genetics, Department of Pediatrics, Mass General for Children, Boston, Massachusetts, USA.

Determining karyotype-phenotype correlations for individuals with Turner syndrome ("TS individuals") is a longstanding research endeavor. The limited literature on Turner syndrome (TS) with a ring X chromosome hinders counseling about the neuropsychological and clinical features. To further characterize these phenotypes, we compared 27 TS individuals with 46,X,r(X)/45,X ("ring X") to 50 non-mosaic 45,X, and 27 mosaic 45,X/46,XX ("mosaic 45,X") individuals.

View Article and Find Full Text PDF

Background: The human microbiome is crucial in regulating intestinal and systemic functions. While its role in cardiovascular disease is better understood, the link between intestinal microbiota and valvular heart diseases (VHD) remains largely unexplored.

Methods: Peer-reviewed studies on human, animal or cell models analysing gut microbiota profiles published up to April 2024 were included.

View Article and Find Full Text PDF

Severe mitral regurgitation (MR) following surgical repair of the mitral valve poses a significant clinical challenge. Patients who have undergone surgery are typically at high risk for a second operation. This report details the case of a 54-year-old male who underwent aortic valve replacement and mitral valve repair using a 34-ring, 14 years prior.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!