Fat-referenced MR thermometry in the breast and prostate using IDEAL.

J Magn Reson Imaging

GE Global Research, 1 Research Circle, Niskayuna, NY 12309, USA.

Published: September 2012

Purpose: To demonstrate a three-echo fat-referenced MR thermometry technique that estimates and corrects for time-varying phase disturbances in heterogeneous tissues.

Materials And Methods: Fat protons do not exhibit a temperature-dependent frequency shift. Fat-referenced thermometry methods exploit this insensitivity and use the signal from fat to measure and correct for magnetic field disturbances. In this study, we present a fat-referenced method that uses interpolation of the fat signal to correct for phase disturbances in fat free regions. Phantom and ex vivo tissue cool-down experiments were performed to evaluate the accuracy of this method in the absence of motion. Non-heated in vivo imaging of the breast and prostate was performed to demonstrate measurement robustness in the presence of systemic and motion-induced field disturbances. Measurement accuracy of the method was compared to conventional proton resonance frequency shift MR thermometry.

Results: In the ex vivo porcine tissue experiment, maximum measurement error of the fat-referenced method was reduced 42% from 3.3 to 1.9°C when compared to conventional MR thermometry. In the breasts, measurement errors were reduced by up to 70% from 6.4 to 1.9°C.

Conclusion: Ex vivo and in vivo results show that the proposed method reduces measurement errors in the heterogeneous tissue experiments when compared to conventional MR thermometry.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jmri.23692DOI Listing

Publication Analysis

Top Keywords

fat-referenced thermometry
12
compared conventional
12
breast prostate
8
phase disturbances
8
frequency shift
8
field disturbances
8
fat-referenced method
8
accuracy method
8
conventional thermometry
8
measurement errors
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!