Recent thoracic artificial lung (TAL) prototypes have impedances lower than the natural lung. With these devices, proximal pulmonary artery (PA) to distal PA TAL attachment may be possible in patients without right ventricular dysfunction. This study examined the relationship between pulmonary system impedance and cardiac output (CO) to create TAL design constraints. A circuit with adjustable resistance and compliance (C) was attached in a PA-PA fashion with the pulmonary circulation of seven sheep with chronic pulmonary hypertension. The pulmonary system zeroth harmonic impedance modulus (Z(0)) was increased by 1, 2.5, and 4 mmHg/(L/min) above baseline. At each Z(0), C was set to 0, 0.34, and 2.1 ml/mmHg. The change in pulmonary system zeroth and first harmonic impedance moduli (ΔZ(0) and ΔZ(1)), the percent change in CO (%ΔCO), and the inlet and outlet anastomoses resistances were calculated for each situation. Results indicate that ΔZ(0) (p < 0.001) but not ΔZ(1) (p = 0.5) had a significant effect on %ΔCO and that %ΔCO = -7.45*ΔZ(0) (R(2) = 0.57). Inlet and outlet anastomoses resistances averaged 0.77 ± 0.16 and 0.10 ± 0.19 mmHg/(L/min), respectively, and the relationship between %ΔCO and TAL resistance, R(T), in mmHg/(L/min) was determined to be %ΔCO = -(7.45f)×(R(T) + 0.87), in which f = the fraction of CO through the TAL. Thus, newer TAL designs can limit %ΔCO to less than 10% if f < 0.75.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3604742PMC
http://dx.doi.org/10.1097/MAT.0b013e318256bb36DOI Listing

Publication Analysis

Top Keywords

pulmonary system
12
thoracic artificial
8
artificial lung
8
pulmonary hypertension
8
system zeroth
8
zeroth harmonic
8
harmonic impedance
8
inlet outlet
8
outlet anastomoses
8
anastomoses resistances
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!