We report a rare complication of spontaneous late dissociation of the tibial insert 2 years after total knee arthroplasty using a high-flex posterior-stabilized Genesis II prosthesis (Smith & Nephew, Memphis, Tenn). It appears that 2 factors may have contributed to dissociation, namely, incomplete seating of the insert and the design of the prosthesis, which involves a shallow anterior tab snap-fit locking mechanism and thin dovetail lips.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.arth.2012.03.056 | DOI Listing |