Objective: To evaluate the agreement of optic disc measurements obtained with the Cirrus high-density optical coherence tomography (HD-OCT) and the Heidelberg retina tomograph (HRT) and compare the intervisit, test-retest variability between the instruments.

Design: Prospective, cross-sectional study.

Participants: Two hundred seven subjects (109 glaucoma and 98 normal subjects).

Methods: One eye from each individual was selected randomly for optic disc imaging by the Cirrus HD-OCT and the HRT. Areas of the optic disc and the cup, cup volume, vertical cup-to-disc ratio and cup-to-disc area ratio were compared between the instruments. The OCT measurements were corrected for ocular magnification using the Littman's formula. The measurement agreement was evaluated with the Bland-Altman plots. The intervisit test-retest variability was examined in 17 randomly selected glaucoma patients who underwent optic disc imaging weekly for 8 consecutive weeks. The intraclass correlation coefficients (ICC) and the reproducibility coefficients of the optic disc parameters were computed.

Main Outcome Measures: Measurement agreement, reproducibility coefficients, and ICCs of optic disc parameters.

Results: The OCT measured smaller optic disc and rim areas and greater cup volume, vertical cup-to-disc ratio and cup-to-disc area ratio than the HRT did (all with P<0.001). There were proportional biases in the Bland-Altman plots between OCT and HRT optic disc measurements except for rim area and cup-to-disc area ratio. The 95% limits of agreement of rim area ranged between -0.28 and 0.88 mm(2) before, and between -0.22 and 0.92 mm(2) after correction for ocular magnification. Both OCT and HRT showed high test-retest reproducibility with ICCs ≥ 0.921. Although the reproducibility coefficient of OCT rim area (0.093 mm(2); 95% confidence interval [CI], 0.081-0.105 mm(2)) was significantly smaller than that of the HRT (0.186 mm(2); 95% CI, 0.163-0.210 mm(2); P = .018), there were no differences in the ICCs between the instruments.

Conclusions: Optic disc assessment by spectral-domain OCT and confocal scanning laser ophthalmoscopy demonstrates poor agreement but similarly low test-retest variability. The source of their disagreement and its effects on the detection of progression require further study.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ophtha.2012.02.033DOI Listing

Publication Analysis

Top Keywords

optic disc
32
disc imaging
12
optic
8
optical coherence
8
coherence tomography
8
intervisit test-retest
8
test-retest variability
8
cup volume
8
volume vertical
8
vertical cup-to-disc
8

Similar Publications

Compartmental analysis of retinal vascular parameters and thickness in myopic eyes using SS-OCTA.

Front Med (Lausanne)

December 2024

Chongqing Key Laboratory of Prevention and Treatment on Major Blinding Diseases, Chongqing Eye Institute, Chongqing Branch (Municipality Division) of National Clinical Research Center for Ocular Diseases, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China.

Background: This study aimed to comprehensively explore the thickness and topographic distributions of retinal vessel alterations of different myopic eyes by using swept-source OCT angiography (SS-OCTA).

Methods: One hundred myopes were included in this observational cross-sectional study. All participants underwent a series of ocular examinations of biometrical parameters, including spherical equivalent refraction (SER), axial length (AL), intraocular pressure (IOP), curvature radius (CR), and others.

View Article and Find Full Text PDF

Terson Syndrome in Two Infants: Case Report and Literature Review.

Cureus

December 2024

Department of Ophthalmology, King Abdulaziz Medical City, National Guard Health Affairs, Riyadh, SAU.

We report the presentation and outcome of Terson syndrome in four eyes of two infants in a tertiary hospital in Saudi Arabia. This is a retrospective report of two infants with Terson syndrome due to accidental traumatic head injuries. Intraoperative screenshots of the posterior pole were taken for both cases.

View Article and Find Full Text PDF

Purpose: Autologous retinal transplantation has been successfully employed in the treatment of large and myopic macular holes that are refractory to standard surgical treatments. Patients transplanted with a peripheral neurosensory retinal graft have shown unexpected improvements in visual acuity. The study aims to investigate if neural integration of the graft takes place in a porcine model of retinal hole.

View Article and Find Full Text PDF

Background: To evaluate the 6-year physiological rates-of-change in ganglion cell inner plexiform layer (GCIPL) and retinal nerve fibre layer (RNFL) thickness measured with optical coherence tomography.

Methods: We included 2202 out of 2661 subjects from the population-based Singapore Chinese Eye Study who returned for follow-up 6 years after baseline examination (follow-up rate 87.7%).

View Article and Find Full Text PDF

General retinal layer segmentation in OCT images via reinforcement constraint.

Comput Med Imaging Graph

December 2024

Beijing Institute of Technology, No. 5, Zhong Guan Cun South Street, Beijing, 100081, China. Electronic address:

The change of layer thickness of retina is closely associated with the development of ocular diseases such as glaucoma and optic disc drusen. Optical coherence tomography (OCT) is a widely used technology to visualize the lamellar structures of retina. Accurate segmentation of retinal lamellar structures is crucial for diagnosis, treatment, and related research of ocular diseases.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!