Purpose: The authors present a calibration method for a prototype proton computed tomography (pCT) scanner. The accuracy of these measurements depends upon careful calibration of the energy detector used to measure the residual energy of the protons that passed through the object.

Methods: A prototype pCT scanner with a cesium iodide (CsI(Tl)) crystal calorimeter was calibrated by measuring the calorimeter response for protons of 200 and 100 MeV initial energies undergoing degradation in polystyrene plates of known thickness and relative stopping power (RSP) with respect to water. Calibration curves for the two proton energies were obtained by fitting a second-degree polynomial to the water-equivalent path length versus calorimeter response data. Using the 100 MeV calibration curve, the RSP values for a variety of tissue-equivalent materials were measured and compared to values obtained from a standard depth-dose range shift measurement using a water-tank. A cylindrical water phantom was scanned with 200 MeV protons and its RSP distribution was reconstructed using the 200 MeV calibration.

Results: It is shown that this calibration method produces measured RSP values of various tissue-equivalent materials that agree to within 0.5% of values obtained using an established water-tank method. The mean RSP value of the water phantom reconstruction was found to be 0.995 ± 0.006.

Conclusions: The method presented provides a simple and reliable procedure for calibration of a pCT scanner.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3338592PMC
http://dx.doi.org/10.1118/1.3700173DOI Listing

Publication Analysis

Top Keywords

pct scanner
12
water-equivalent path
8
path length
8
prototype proton
8
calibration method
8
calorimeter response
8
100 mev
8
rsp values
8
tissue-equivalent materials
8
water phantom
8

Similar Publications

Objective: Customized nonabsorbable membranes are widely used in severe alveolar bone defects and provide sufficient and precise regenerated bone tissue for subsequent dental implant placement. Although 3D-printed polyetheretherketone (PEEK) meshes have confirmed successful use in clinical cases, the performance of a PEEK mesh is not satisfactory. Compared with PEEK, polyetherketoneketone (PEKK) has better mechanical and processing properties.

View Article and Find Full Text PDF

Accuracy and longitudinal consistency of PET/MR attenuation correction in amyloid PET imaging amid software and hardware upgrades.

AJNR Am J Neuroradiol

September 2024

From Mallinckrodt Institute of Radiology (Chunwei Ying, Shaney Flores, Tammie L. S. Benzinger, Hongyu An), Department of Neurology (Yasheng Chen, Hongyu An), Department of Surgery (Yan Yan), Knight Alzheimer Disease Research Center (Tammie L. S. Benzinger), and Department of Neurosurgery (Tammie L. S. Benzinger), Washington University School of Medicine, St. Louis, Missouri, USA.

Background And Purpose: Integrated PET/MR allows the simultaneous acquisition of PET biomarkers and structural and functional MRI to study Alzheimer disease (AD). Attenuation correction (AC), crucial for PET quantification, can be performed using a deep learning approach, DL-Dixon, based on standard Dixon images. Longitudinal amyloid PET imaging, which provides important information about disease progression or treatment responses in AD, is usually acquired over several years.

View Article and Find Full Text PDF

Proton therapy administers a highly conformal dose to the tumour region, necessitating accurate prediction of the patient's 3D map of proton relative stopping power (RSP) compared to water. This remains challenging due to inaccuracies inherent in single-energy computed tomography (SECT) calibration. Recent advancements in spectral x-ray CT (xCT) and proton CT (pCT) have shown improved RSP estimation compared to traditional SECT methods.

View Article and Find Full Text PDF

Noninvasive quantification of [F]SynVesT-1 binding using simplified reference tissue model 2.

Eur J Nucl Med Mol Imaging

December 2024

Department of Radiology and Biomedical Imaging, Yale School of Medicine, 801 Howard Avenue, PO Box 208048, New Haven, CT, 06520- 8048, USA.

Purpose: [F]SynVesT-1, a positron emission tomography (PET) radiotracer for the synaptic vesicle glycoprotein 2A (SV2A), demonstrates kinetics similar to [C]UCB-J, with high brain uptake, fast kinetics fitting well with the one-tissue compartment (1TC) model, and excellent test-retest reproducibility. Challenges arise due to the similarity between k and [Formula: see text] (efflux rate of the reference region), when applying the simplified reference tissue model (SRTM) and related methods in [C]UCB-J studies to accurately estimate [Formula: see text]. This study evaluated the suitability of these methods to estimate [F]SynVesT-1 binding using centrum semiovale (CS) or cerebellum (CER) as reference regions.

View Article and Find Full Text PDF

Incomplete Hippocampal Inversion (IHI), sometimes called hippocampal malrotation, is an atypical anatomical pattern of the hippocampus found in about 20% of the general population. IHI can be visually assessed on coronal slices of T1 weighted MR images, using a composite score that combines four anatomical criteria. IHI has been associated with several brain disorders (epilepsy, schizophrenia).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!