Evaluation of total hip arthroplasty devices using a total joint replacement registry.

Pharmacoepidemiol Drug Saf

Department of Surgical Outcomes and Analysis, Kaiser Permanente, San Diego, CA 92109, USA.

Published: May 2012

Purpose: The purpose of this paper is to describe the infrastructure of the total joint replacement registry of a large integrated healthcare system's and emphasize challenges associated with orthopedic device classification and evaluation.

Methods: Using a large integrated healthcare system innovative infrastructure including electronic health record data, administrative data sources, and registry data collection, we evaluated device choice and outcomes of total hip arthroplasty (THA). Devices were classified into type of bearing surface (alternative versus traditional). Multiple imputation was used to accommodate missing data, and a logistic regression model was applied to assess the impact of patient and surgeon factors on choice of bearing surface. A Cox regression model was used to evaluate risk of aseptic revision while controlling for surgeon, site, and patient characteristics. Adjusted cumulative probability-of-event curves were created, comparing survival of alternative against traditional bearings of devices, with aseptic revision as the outcome of interest.

Results: The study sample consisted of 25,377 primary THAs with an average follow-up of 2.7 years. Choice of bearing surface varied by surgeon and patient characteristics. After adjusting for patient, surgeon, and hospital covariates, results showed that the risk of aseptic revision associated with alternative bearings did not differ significantly from traditional bearing surfaces (hazard ratio = 1.33; 95% confidence interval: 0.90, 1.98).

Conclusions: Clinically rich data from a registry with linkages to electronic health records and other administrative databases improve identification of exposures, outcomes, and patient subgroups in medical device evaluation. These various data sources facilitate refined adjustment for potential confounders such as hospital, surgeon, and patient factors and ensure comprehensive device performance evaluation within registries.

Download full-text PDF

Source
http://dx.doi.org/10.1002/pds.3228DOI Listing

Publication Analysis

Top Keywords

bearing surface
12
aseptic revision
12
total hip
8
hip arthroplasty
8
total joint
8
joint replacement
8
replacement registry
8
large integrated
8
integrated healthcare
8
electronic health
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!