Background: This study aimed to quantify the clinical parameters of mono- and bipolar instruments that inhibit pacemaker function. The specific aims were to quantify pacer inhibition resulting from the monopolar instrument by altering the generator power setting, the generator mode, the distance between the active electrode and the pacemaker, and the location of the dispersive electrode.

Methods: A transvenous ventricular lead pacemaker overdrive paced the native heart rate of an anesthetized pig. The primary outcome variable was pacer inhibition quantified as the number of beats dropped by the pacemaker during 5 s of monopolar active electrode activation.

Results: Lowering the generator power setting from 60 to 30 W decreased the number of dropped paced events (2.3 ± 1.2 vs 1.6 ± 0.8 beats; p = 0.045). At 30 W of power, use of the cut mode decreased the number of dropped paced beats compared with the coagulation mode (0.6 ± 0.5 vs 1.6 ± 0.8; p = 0.015). At 30 W coagulation, firing the active electrode at different distances from the pacemaker generator (3.75, 7.5, 15, and 30 cm) did not change the number of dropped paced beats (p = 0.314, analysis of variance [ANOVA]). The dispersive electrode was placed in four locations (right/left gluteus, right/left shoulder). More paced beats were dropped when the current vector traveled through the pacemaker/leads than when it did not (1.5 ± 1.0 vs 0.2 ± 0.4; p < 0.001).

Conclusions: Clinical parameters that reduce the inhibition of a pacemaker by monopolar instruments include lowering the generator power setting, using cut (vs coagulation) mode, and locating the dispersive electrode so the current vector does not traverse the pacemaker generator or leads.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00464-012-2279-3DOI Listing

Publication Analysis

Top Keywords

generator power
12
power setting
12
active electrode
12
number dropped
12
dropped paced
12
paced beats
12
pacemaker
8
pacemaker function
8
clinical parameters
8
pacer inhibition
8

Similar Publications

Introduction: Women have generally lower body size and lean- to fat-mass ratio, lower maximal anaerobic power due to a lower muscle mass, and fewer fast-twitch fibers, although they can show higher resistance to fatigue or greater metabolic flexibility than men. These factors are well known and explain the sex differences in endurance sports such as distance running (10%-12%). Several of these factors-particularly the differences in body composition and skeletal-muscle characteristics-may directly impact vertical displacement and uphill performance.

View Article and Find Full Text PDF

Machine learning techniques for the prediction of indoor gamma-ray dose rates - Strengths, weaknesses and implications for epidemiology.

J Environ Radioact

December 2024

Radiation Epidemiology Branch, Division of Cancer Epidemiology and Genetics, National Cancer Institute, DHHS, NIH, Bethesda, MD, 20892-9778, USA; Faculty of Health and Life Sciences, Oxford Brookes University, Headington Campus, Oxford, OX3 0BP, UK. Electronic address:

We investigate methods that improve the estimation of indoor gamma ray dose rates at locations where measurements had not been made. These new predictions use a greater range of modelling techniques and larger variety of explanatory variables than our previous examinations of this subject. Specifically, we now employ three types of machine learning models in addition to the geostatistical, nearest neighbour and other earlier models.

View Article and Find Full Text PDF

Background: Patients with type 1 diabetes (DM1), even in the setting of adequate glycaemic control, have an excess risk for developing cardiovascular disease. Residual insulin secretion (RIS), measured by detectable C-peptide levels in patients with DM1, might protect against diabetes-related complications. This study aimed to examine the relationship between residual insulin secretion and prognostic markers of cardiovascular complications in patients with DM1.

View Article and Find Full Text PDF

Dual-compartment-gate organic transistors for monitoring biogenic amines from food.

Biosens Bioelectron

December 2024

Department of Life Sciences, Università Degli Studi di Modena e Reggio Emilia, Via Campi 103, Modena, 41125, Italy; Center for Translational Neurophysiology of Speech and Communication, Istituto Italiano di Tecnologia (CTNSC), Via Fossato di Mortara 17-19, Ferrara, 44121, Italy.

According to the Food and Agriculture Organization of the United Nations (FAO) more than 14% of the world's food production is lost every year before reaching retail, and another 17% is lost during the retail stage. The use of the expiration date as the main estimator of the life-end of food products creates unjustified food waste. Sensors capable of quantifying the effective food freshness and quality could substantially reduce food waste and enable more effective management of the food chain.

View Article and Find Full Text PDF

Objectives: To investigate maxillary canine movement accuracy and anchorage during space closure in first premolar extraction cases (maximum anchorage) using In-House Clear Aligners (IHCAs).

Materials And Methods: A randomised controlled trial with a split-mouth design recruited 16 adults in university setting. Each patient was randomly assigned by side for canine retraction using 12 IHCAs to both the experimental palatal power arm (Pa) and non-Pa control (C).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!