Electrical desynchronization in cardiac resynchronization therapy (CRT) occurs when sinus P waves are continually locked in the postventricular atrial refractory period (PVARP). This process is characterized by sequences of a P wave as an atrial event in the PVARP followed by a conducted and sensed ventricular event. Such sequences are more common in patients with a prolonged PR interval, often initiated by premature ventricular complexes (PVC) and terminated by PVCs or slowing of the sinus rate. Specific algorithms automatically identify a recurring pattern of P wave locking in the PVARP, whereupon they shorten the PVARP temporarily until atrial tracking is restored with the programmed sensed AV interval. The Biotronik family of Lumax CRT devices use an AV control window which is not an algorithm that "unlocks" P waves trapped in the PVARP. Rather, it prevents P waves from becoming trapped in the PVARP. A ventricular sensed event occurring within the AV control interval does not start a PVARP so that P wave locking cannot occur when the AV conduction time is shorter than the AV control interval.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00399-012-0173-2DOI Listing

Publication Analysis

Top Keywords

postventricular atrial
8
atrial refractory
8
refractory period
8
cardiac resynchronization
8
wave locking
8
waves trapped
8
trapped pvarp
8
control interval
8
pvarp
7
p-wave locking
4

Similar Publications

Article Synopsis
  • The study explores a new type of heart rhythm issue called pseudo-RNRVAS that occurs when atrial stimulation stops while there is continued communication between the ventricles and atrium during certain heart phases.
  • Researchers investigated 21 patients with cardiac devices, finding that 20 experienced pseudo-RNRVAS, particularly at lower heart rates and specific timing intervals after losing atrial capture.
  • The findings suggest that pseudo-RNRVAS can happen without the usual risk factors and are more likely in scenarios that increase atrial pacing, especially soon after pacemaker implantation, indicating a possible link to pacemaker syndrome and heart failure risk.
View Article and Find Full Text PDF
Article Synopsis
  • Endless loop tachycardia (ELT) is a frequent issue in patients with implanted cardiac devices and can cause serious symptoms despite preventive measures.
  • * A study analyzed 20 cases of sustained ELT from 4,520 device checkups over four years, finding it usually benign but occasionally harmful, particularly in patients with left ventricular dysfunction.
  • * Various factors contributed to ELT, including improper AV synchrony and sensing issues, and adjustments like changing sensitivity settings or PVARP duration were necessary for management.
View Article and Find Full Text PDF

2:1 electronic AV block due to inappropriate automatic post‐ventricular atrial refractory period extension.

Pacing Clin Electrophysiol

October 2024

Department of Clinical and Experimental Medicine-Paediatric Cardiology Unit, University of Messina, Messina, Italy.

A 16-year-old female with dual-chamber pacemaker (Medtronic Azure XT DR), due to symptomatic third-degree congenital atrioventricular (AV) block, presented to our ambulatory with dizziness and presyncopal episodes preceded by prodromes, occurring over the last few months. The device was programmed in DDD mode with an upper rate of 150 bpm. A head-up Tilt Test (HUTT) revealed the unexpected emergence of 2:1 electronic AV block at a sinus rate of 130 bpm.

View Article and Find Full Text PDF

An automatic pacemaker algorithm causing exertional intolerance.

J Cardiovasc Electrophysiol

November 2023

Department of Medicine, Division of Cardiology, Weill Cornell Medicine, New York Presbyterian Hospital, New York, New York, USA.

View Article and Find Full Text PDF

Background: We recently described a novel pacemaker-mediated arrhythmia in Abbott cardiac implantable electronic devices (CIED), called pseudo-RNRVAS, that mimics repetitive non-reentrant ventriculoatrial synchrony (RNRVAS), but can appear in patients with ventriculoatrial (VA) block. It is caused by sinus-p-waves, trapped in the post-ventricular atrial refractory period (PVARP), which mimic VA conduction. The p-waves are followed by atrial pacing during the myocardial refractory time, which can trigger atrial fibrillation (AF).