In vitro ultrasound characterization of a polyurethane trileaflet valve.

ASAIO Trans

Cardiac Surgery Branch, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892.

Published: January 1991

Polyurethane synthetic trileaflet valves were compared with commercial prostheses in vitro, in a pulse duplicator using ultrasound to characterize the flow velocities and patterns. Flow-pressure drop behavior was in the middle range of other prosthetic valves. Diastolic regurgitant jets were located by color Doppler ultrasound, and there appeared to be some leakage through the leaflet fold at the commissure. Nevertheless, the closing volumes and closed valve leakage volumes were, on average, lower than other prosthetic valves. Systolic 20 Hz spectral oscillations detected with pulsed Doppler and continuous wave Doppler were attributed to leaflet flutter in the open valve.

Download full-text PDF

Source

Publication Analysis

Top Keywords

prosthetic valves
8
vitro ultrasound
4
ultrasound characterization
4
characterization polyurethane
4
polyurethane trileaflet
4
trileaflet valve
4
valve polyurethane
4
polyurethane synthetic
4
synthetic trileaflet
4
trileaflet valves
4

Similar Publications

Background: The initial outcomes of transcatheter aortic valve replacement in patients with left ventricular outflow tract calcification are poor. Furthermore, balloon-expandable transcatheter aortic valve replacement is associated with an increased risk of annular rupture, and self-expandable transcatheter aortic valve replacement is associated with worse post-operative residual paravalvular leakage grades. Therefore, developing an optimal method for transcatheter aortic valve replacement for patients with left ventricular outflow tract calcification is desirable.

View Article and Find Full Text PDF

Background: Patients after transcatheter pulmonary valve implantation (TPVI) are at increased risk for infective prosthetic valve endocarditis. Diagnosis of infective endocarditis (IE) following TPVI is particularly difficult due to impaired visualization of the transcatheter pulmonary valve (TPV) with echocardiography [Delgado V, Ajmone Marsan N, de Waha S, Bonaros N, Brida M, Burri H, et al. 2023 ESC guidelines for the management of endocarditis.

View Article and Find Full Text PDF

Background: Since the transcatheter valve-in-valve (ViV) procedure was introduced in 2007, a few cases of infective endocarditis (IE) following the ViV procedure have been reported, which can be predisposed by older age, pre-existing medical conditions, and procedural techniques. Paravalvular abscesses constitute a rare complication of IE, resulting from extending IE beyond the valve annulus, less commonly caused by species. This complication is more common in prosthetic valves, particularly bioprosthetic valves.

View Article and Find Full Text PDF

Background: Transcatheter valve-in-valve replacement (TMViVR) is an alternative option for patients with bioprosthetic valve failure (BVF) who are at high surgical risk. Although infective endocarditis (IE) after transcatheter mitral valve-in-valve replacement is unusual, it is associated with significantly high mortality.

Case Presentation: An 81-year-old male patient was admitted with intermittent thoracic tightness, chest pain persisting for 3 years, and shortness of breath with nausea for 1 week.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!