Objective: Although the intra-aortic balloon pump is the most used ventricular assist device, no study has ever evaluated the best weaning method. We compared 2 different intra-aortic balloon pump weaning methods.

Methods: Thirty consecutive patients needing an intra-aortic balloon pump because of perioperative low-output cardiac syndrome were randomized to be weaned by ratio (4 consecutive hours of a 1:2 assisting ratio followed by 1 hour of a 1:3 ratio; group R) or by progressive volume deflation (10% of total volume every hour for 5 consecutive hours; 15 patients, group V). A duration of 5 hours was set a priori as the weaning duration. The weaning protocol was started when the cardiac index was greater than 2.5 L/min/m(2), the central venous pressure was 12 mm Hg or less, the blood lactate was less than 2.5 mmol/L, the mean arterial pressure was greater than 65 mm Hg, and the preserved urine output (≥1 mL/kg/hr) lasted for at least 5 consecutive hours before weaning. The cardiac index, indexed systemic vascular resistance, cardiac cycle efficiency, and central venous pressure were registered at 9 points (T0, start; T1 to T5, the first 5 weaning hours; T6, 2 hours after withdrawal; T7, 12 hours after withdrawal; and T8, at intensive care unit discharge) using the pressure recording analytical method. The interval from intra-aortic balloon pump withdrawal to intensive care unit discharge, weaning failure, perioperative troponin I, and lactate (same points) were compared.

Results: All patients, except for 1 belonging to group R (P = 1.0), were successfully weaned. Group V had better preserved cardiac index, indexed systemic vascular resistance, cardiac cycle efficiency, and central venous pressure (group*time P = .0001). Group R had worse cardiac index from T5 to T8 (P ≤ .0001), indexed systemic vascular resistance from T2 to T8 (P ≤ .004), cardiac cycle efficiency from T3 to T8 (P ≤ .001), central venous pressure from T4 to T8 (P ≤ .0001), and a longer interval from intra-aortic balloon pump withdrawal to intensive care unit discharge (P = .0001). The lactate level was lower in group V from T5 to T8 (P ≤ .027; group*time P = .001).

Conclusions: Intra-aortic balloon pump weaning by volume deflation allowed better hemodynamic and metabolic parameters.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jtcvs.2012.03.063DOI Listing

Publication Analysis

Top Keywords

intra-aortic balloon
28
balloon pump
24
central venous
16
venous pressure
16
consecutive hours
12
indexed systemic
12
systemic vascular
12
vascular resistance
12
cardiac cycle
12
cycle efficiency
12

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!