Objectives: We investigate the influence of stent design on stent coverage at 6-9 months after sirolimus eluting stent (SES) implantation using optical coherence tomography (OCT).

Background: Although some reports suggest that stent design may correlate with stent coverage of stent struts, there were few detailed data whether stent design impact on stent coverage in the same drug-eluting stent.

Methods: A total of 21 SESs (15 patients), who had implanted 2.5, 2.75, and 3.0 mm stents, underwent OCT at 6-9 months after stent implantation. SES is constructed by two different strut width-components; narrow strut width parts (59 μm) and wide strut width parts (115 μm). Thus, we divided stent struts of SESs into two groups: narrow strut width parts (narrow group) and wide ones (wide group). We compared the incidence of incomplete apposed struts, uncovered struts, and neointimal hyperplasia (NIH) thickness between the two groups.

Results: We could detect 2,948 struts (narrow group consisted of 1,132 struts and wide group consisted of 1,816 struts). Incidence of uncovered struts in the narrow group was significantly lower than in the wide group (30.2% vs. 40.8%, P < 0.001), and NIH thickness in the narrow group was significantly greater than in the wide group (127.5 ± 93.4 μm vs. 118.6 ± 81.4 μm, P = 0.03).

Conclusions: Stent design, especially strut width, affects stent coverage of SES. The narrow strut may avoid the absence of stent coverage in SES, which correlates with stent thrombosis.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.24401DOI Listing

Publication Analysis

Top Keywords

stent coverage
24
strut width
20
stent design
16
narrow group
16
wide group
16
stent
15
narrow strut
12
width parts
12
width stent
8
optical coherence
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!