Introduction: Problems with loosening of the modules for the modular endoprosthesis were encountered in animal studies for mandibular body replacement. We performed a finite element analysis to look at the stress distribution and areas of stress concentration in a human sized mandible. Variations were made to the stem and defect length to look at how the forces changed. The hypothesis was: (1) reconstruction with a modular endoprosthesis did not lead to areas of stress concentration beyond the material strength of cortical bone and titanium alloy; (2) changes in dimensions of the endoprosthesis did not cause a corresponding linear increase to the stresses.

Materials And Methods: The endoprosthesis was modelled to create a male, female part with stems and a connection screw (Case I). The stem length was halved (Case II) and defect length doubled (Case III). Geometric data of a human sized mandible were obtained, a continuity defect created digitally at the right molar area and the models combined. Boundary conditions were set and the model loaded to get a bite force of 300 N at the incisor region. An intact mandible was used as a control.

Results: The right side of the reconstructed mandible became less rigid and flexed more. The highest stresses were within the endoprosthesis at two areas of stress concentration: (1) shear stress at the superior surface of the stems close to the junction of the stem and the module body; (2) compressive stresses at the bottom bevel of the dove-tailed connection. The stress distribution for Case I and II did not differ much except for the magnitude which was slightly higher for Case II. There was a tendency for outward bending at the module connection for Case III which potentially might cause loosening of the module connection. Displacements of the mandible were less than 1 mm throughout.

Conclusion: The endoprosthesis with its present dimensions would be expected to perform adequately at a bite force of 300 N. An increase in defect length caused a tendency for bending at the stem and the module connection. With a decrease in stem length, there were little differences except a slight increase in magnitude.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcms.2012.03.010DOI Listing

Publication Analysis

Top Keywords

modular endoprosthesis
12
areas stress
12
stress concentration
12
defect length
12
module connection
12
mandibular body
8
body replacement
8
finite element
8
element analysis
8
stress distribution
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!