Background: Since Atkinson's original description of retrobulbar block in 1936, needle-based anesthetic techniques have become integral to ophthalmic anesthesia. These techniques are unfortunately associated with rare, grave complications such as globe perforation. Ultrasound has gained widespread acceptance for peripheral nerve blockade, but its translation to ocular anesthesia has been hampered because sonic energy, in the guise of thermal or biomechanical insult, is potentially injurious to vulnerable eye tissue. The US Food and Drug Administration (FDA) has defined guidelines for safe use of ultrasound for ophthalmic examination, but most ultrasound devices used by anesthesiologists are not FDA-approved for ocular application because they generate excessive energy. Regulating agencies state that ultrasound examinations can be safely undertaken as long as tissue temperatures do not increase >1.5°C above physiological levels.
Methods: Using a rabbit model, we investigated the thermal and mechanical ocular effects after prolonged ultrasonic exposure to single orbital- and nonorbital-rated devices. In a dual-phase study, aimed at detecting ocular injury, the eyes of 8 rabbits were exposed to continuous 10-minute ultrasound examinations from 2 devices: (1) the Sonosite Micromaxx (nonorbital rated) and (2) the Sonomed VuMax (orbital rated) machines. In phase I, temperatures were continuously monitored via thermocouples implanted within specific eye structures (n = 4). In phase II the eyes were subjected to ultrasonic exposure without surgical intervention (n = 4). All eyes underwent light microscopy examinations, followed at different intervals by histology evaluations conducted by an ophthalmic pathologist.
Results: Temperature changes were monitored in the eyes of 4 rabbits. The nonorbital-rated transducer produced increases in ocular tissue temperature that surpassed the safe limit (increases >1.5°C) in the lens of 3 rabbits (at 5.0, 5.5, and 1.5 minutes) and cornea of 2 rabbits (both at 1.5 minutes). A secondary analysis of temporal temperature differences between the orbital-rated and nonorbital transducers revealed statistically significant differences (Bonferroni-adjusted P < 0.05) in the cornea at 3.5 minutes, the lens at 2.5 minutes, and the vitreous at 4.0 minutes. Light microscopy and histology failed to elicit ocular injury in either group.
Conclusions: The nonorbital-rated ultrasound machine (Sonosite Micromaxx) increases the ocular tissue temperature. A larger study is needed to establish safety. Until then, ophthalmic ultrasound-guided blocks should only be performed with ocular-rated devices.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3381790 | PMC |
http://dx.doi.org/10.1213/ANE.0b013e318253622e | DOI Listing |
Am J Physiol Cell Physiol
January 2025
Laboratoire de Physiopathologie et Régulation des Transports Ioniques, Université de Poitiers, France.
Despite the importance of ocular surface in human physiology and diseases, little is known about ion channel expression, properties and regulation in ocular epithelial cells. Furthermore, human primary epithelial cells have rarely been studied in favor of rat, mouse and especially rabbit animal models. Here, we developed primary human Meibomian gland (hMGEC) and conjunctival (hConEC) epithelial cells.
View Article and Find Full Text PDFHistol Histopathol
January 2025
University of Arizona, Phoenix, AZ, USA.
In larger, translational animal models, manual measurements of longitudinal bone growth using fluorochrome labels is tedious and may be prone to less rigor due to variations in reader experience, sampling differences, and photobleaching that limits the repeatability of measurements. This study assesses the reliability of three different digital methods to assist in measurement of distance between pulsed fluorochrome labels. Forty-five tibial physes from skeletally immature New Zealand White rabbits were pulsed with fluorochrome labels and measured using Fully Manual Technique (FMT), Manual Digital Measurement (MDM), Computer Assisted Image Processing (AIP), and Fully Automated Measurement (FAM).
View Article and Find Full Text PDFInt J Nanomedicine
January 2025
Shanghai Eighth People's Hospital, Xuhui District, Shanghai, 200030, People's Republic of China.
Purpose: Gastrostomy is the commonly used enteral feeding technology. The clinical risks caused by tube dislodgement and peristomal site infection are the common complications before complete tract maturation after gastrostomy. However, there is currently no relevant research to promote gastrostomy wound treatment and tract maturation.
View Article and Find Full Text PDFPeerJ
January 2025
College of Engineering, University of Tennessee - Knoxville, TN, United States of America.
Objective: The purpose of this study was to investigate the timing and mode of failure of metallic screw-type suture anchors used to attach artificial tendons to bone in an New Zealand White rabbit model.
Study Design: Metal suture anchors with braided composite sutures of varying sizes (United States Pharmacopeia (USP) size 1, 2, or 5) were used to secure artificial tendons replacing both the Achilles and tibialis cranialis tendons in 12 female New Zealand White rabbits. Artificial tendons were implanted either at the time of (immediate replacement, = 8), or four/five weeks after (delayed replacement, = 4) resection of the biological tendon.
Adv Healthc Mater
January 2025
Nankai University Eye Institute, Nankai University, Tianjin, 300071, China.
Reproducing the microstructure of the natural cornea remains a significant challenge in achieving the mechanical and biological functionality of artificial corneas. Therefore, the development of cascade structures that mimic the natural extracellular matrix (ECM), achieving both macro-stability and micro-structure, is of critical importance. This study proposes a novel, efficient, and general photo-functionalization strategy for modifying natural biomaterials.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!