The Columbus knee system is designed as a standard knee implant to allow high flexion without additional bone resection. Between August, 2004 and March, 2010 we performed 109 total knee arthroplasties of the Columbus knee system in 101 consecutive patients suffering from primary arthrosis of the knee. Mean age was 72.4 years in women and 70.3 years in men. Mean followup was 47.3 months. The 4-year results of a group of patients who received the NexGen Full Flex implant operated by the same surgeon were used for comparison. Mean total knee score was Columbus: 175.6 and NexGen Flex: 183.4; P = 0.037. Mean operation time was 53 min for Columbus and 66 min for NexGen Flex; P < 0.001. With new streamlined instruments operative time became 60 min for the Columbus; P > 0.05. Radiological assessment showed no signs of loosening for both groups. Therefore, the Columbus knee system can be recommended for flexion angles up to 140°.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3310259 | PMC |
http://dx.doi.org/10.1155/2012/213817 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!