Background: Transthoracic echocardiography (TTE) is the standard method for the evaluation of the severity of aortic stenosis (AS). Valve effective orifice area (EOA) measured by the continuity equation is one of the most frequently used stenotic indices. However, TTE measurement of aortic valve EOA is not feasible or not reliable in a significant proportion of patients. Cardiovascular magnetic resonance (CMR) has emerged as a non-invasive alternative to evaluate EOA using velocity measurements. The objectives of this study were: 1) to validate a new CMR method using jet shear layer detection (JSLD) based on acoustical source term (AST) concept to estimate the valve EOA; 2) to introduce a simplified JSLD method not requiring vorticity field derivation.

Methods And Results: We performed an in vitro study where EOA was measured by CMR in 4 fixed stenoses (EOA = 0.48, 1.00, 1.38 and 2.11 cm²) under the same steady flow conditions (4-20 L/min). The in vivo study included eight (8) healthy subjects and 37 patients with mild to severe AS (0.72 cm² ≤ EOA ≤ 1.71 cm²). All subjects underwent TTE and CMR examinations. EOA was determinated by TTE with the use of continuity equation method (TTE(CONT)). For CMR estimation of EOA, we used 3 methods: 1) Continuity equation (CMR(CONT)); 2) Shear layer detection (CMR(JSLD)), which was computed from the velocity field of a single CMR velocity profile at the peak systolic phase; 3) Single plane velocity truncation (CMR(SPVT)), which is a simplified version of CMR(JSLD) method. There was a good agreement between the EOAs obtained in vitro by the different CMR methods and the EOA predicted from the potential flow theory. In the in vivo study, there was good correlation and concordance between the EOA measured by the TTE(CONT) method versus those measured by each of the CMR methods: CMR(CONT) (r = 0.88), CMR(JSLD) (r = 0.93) and CMR(SPVT) (r = 0.93). The intra- and inter- observer variability of EOA measurements was 5 ± 5% and 9 ± 5% for TTE(CONT), 2 ± 1% and 7 ± 5% for CMR(CONT), 7 ± 5% and 8 ± 7% for CMR(JSLD), 1 ± 2% and 3 ± 2% for CMR(SPVT). When repeating image acquisition, reproducibility of measurements was 10 ± 8% and 12 ± 5% for TTE(CONT), 9 ± 9% and 8 ± 8% for CMR(CONT), 6 ± 5% and 7 ± 4% for CMR(JSLD) and 3 ± 2% and 2 ± 2% for CMR(SPVT).

Conclusion: There was an excellent agreement between the EOA estimated by the CMR(JSLD) or CMR(SPVT) methods and: 1) the theoretical EOA in vitro, and 2) the TTE(CONT) EOA in vivo. The CMR(SPVT) method was superior to the TTE and other CMR methods in terms of measurement variability. The novel CMR-based methods proposed in this study may be helpful to corroborate stenosis severity in patients for whom Doppler-echocardiography exam is inconclusive.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3366866PMC
http://dx.doi.org/10.1186/1532-429X-14-23DOI Listing

Publication Analysis

Top Keywords

eoa
15
eoa measured
12
continuity equation
12
cmr methods
12
cmr
9
cardiovascular magnetic
8
magnetic resonance
8
aortic stenosis
8
stenosis severity
8
single plane
8

Similar Publications

Background: Cementless stems are commonly used in hemiarthroplasty (HA) for femoral neck fractures. Recent studies have reported increased risk of periprosthetic fracture with cementless stems compared to cemented HA. In elective total hip arthroplasty (THA), lower proximal canal fill ratios (CFR) of cementless stems have been associated with worse outcomes.

View Article and Find Full Text PDF

Background: The management choice for the middle cerebral artery aneurysms (MCAAs) is still controversial. This review aims to describe a single-center "clipping first" policy for MCAA over 40 years of experience and compare the short- and long-term clinical outcomes by aneurysm's location.

Methods: This retrospective cohort study reviews the whole series of a single-center intracranial aneurysm mainly based on the micro-neurosurgical experience of the senior authors (EOA and EKA).

View Article and Find Full Text PDF

It is urgent to develop effective antibiotic alternatives for the control of subclinical necrotic enteritis (NE) in chickens after in-feed antibiotics have been banned. The current study investigated the efficacy of drinking water supplemented with essential oils and organic acids mixtures (EOA) on growth performance and intestinal health of broilers challenged with necrotic enteritis (NE). A total of 360 one-day-old Arbor Acres male broilers were randomly divided into 5 treatment groups, including non-challenged control group (T0), challenged NE group (T1), and challenged NE chickens treated with 0.

View Article and Find Full Text PDF

Assessing the basis for regulatory crediting of virus LRVs for secondary biological wastewater treatment: A systematic review.

Water Res

March 2025

Southern Nevada Water Authority, P.O. Box 99954, Las Vegas, NV 89193, United States. Electronic address:

Regulatory frameworks for potable reuse often include stringent log reduction value (LRV) targets to ensure public health protection against exposure to viruses and protozoa. To achieve overall LRV targets and reduce associated capital and operational costs, it is important to maximize LRV credits awarded to each unit process in a potable reuse treatment train. This may include processes that are historically uncredited or undercredited, such as secondary biological wastewater treatment incorporating activated sludge and secondary clarification.

View Article and Find Full Text PDF

Impact of Measured and Predicted Prosthesis-Patient Mismatch After Transcatheter Aortic Valve Replacement.

JACC Cardiovasc Interv

November 2024

University of Cologne, Faculty of Medicine and University Hospital Cologne, Clinic III for Internal Medicine, Cologne, Germany. Electronic address:

Background: Prosthesis-patient mismatch after transcatheter aortic valve replacement (TAVR) can be measured echocardiographically (measured prosthesis-patient mismatch [PPMm]) or predicted (predicted prosthesis-patient mismatch [PPMp]) using published effective orifice area (EOA) reference values. However, the clinical implications of PPM post-TAVR remain unclear.

Objectives: This study aimed to elucidate the prevalence of PPMm and PPMp post-TAVR and their impact on mortality in a large international cohort.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!