Automated methods for accurate determination of the critical velocity of packed bed chromatography.

Biotechnol Prog

The Advanced Centre for Biochemical Engineering, Department of Biochemical Engineering, University College London, London WC1E 7JE, UK.

Published: November 2012

Knowing the critical velocity (ucrit) of a chromatography column is an important part of process development as it allows the optimization of chromatographic flow conditions. The conventional flow step method for determining ucrit is prone to error as it depends heavily on human judgment. In this study, two automated methods for determining ucrit have been developed: the automatic flow step (AFS) method and the automatic pressure step (APS) method. In the AFS method, the column pressure drop is monitored upon application of automated incremental increases in flow velocity, whereas in the APS method the flow velocity is monitored upon application of automated incremental increases in pressure drop. The APS method emerged as the one with the higher levels of accuracy, efficiency and ease of application having the greater potential to assist defining the best operational parameters of a chromatography column.

Download full-text PDF

Source
http://dx.doi.org/10.1002/btpr.1544DOI Listing

Publication Analysis

Top Keywords

aps method
12
automated methods
8
critical velocity
8
chromatography column
8
flow step
8
determining ucrit
8
afs method
8
pressure drop
8
monitored application
8
application automated
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!