Assessment of quantitative perfusion parameters by dynamic contrast-enhanced sonography using a deconvolution method: an in vitro and in vivo study.

J Ultrasound Med

Laboratoire d'Imagerie du Petit Animal, Unité Mixte de Recherche, Institut Gustave Roussy, Pavillon de Recherche I, 39 rue Camille Desmoulins, 94805 Villejuif, France.

Published: April 2012

Objectives: The purpose of this study was to investigate the impact of the arterial input on perfusion parameters measured using dynamic contrast-enhanced sonography combined with a deconvolution method after bolus injections of a contrast agent.

Methods: The in vitro experiments were conducted using a custom-made setup consisting of pumping a fluid through a phantom made of 3 intertwined silicone pipes, mimicking a complex structure akin to that of vessels in a tumor, combined with their feeding pipe, mimicking the arterial input. In the in vivo experiments, B16F10 melanoma cells were xenografted to 5 nude mice. An ultrasound scanner combined with a linear transducer was used to perform pulse inversion imaging based on linear raw data throughout the experiments. A mathematical model developed by the Gustave Roussy Institute (patent WO/2008/053268) and based on the dye dilution theory was used to evaluate 7 semiquantitative perfusion parameters directly from time-intensity curves and 3 quantitative perfusion parameters from the residue function obtained after a deconvolution process developed in our laboratory based on the Tikhonov regularization method. We evaluated and compared the intraoperator variability values of perfusion parameters determined after these two signal-processing methods.

Results: In vitro, semiquantitative perfusion parameters exhibited intraoperator variability values ranging from 3.39% to 13.60%. Quantitative parameters derived after the deconvolution process ranged from 4.46% to 11.82%. In vivo, tumors exhibited perfusion parameter intraoperator variability values ranging from 3.74% to 29.34%, whereas quantitative ones varied from 5.00% to 12.43%.

Conclusions: Taking into account the arterial input in evaluating perfusion parameters improves the intraoperator variability and may improve the dynamic contrast-enhanced sonographic technique.

Download full-text PDF

Source
http://dx.doi.org/10.7863/jum.2012.31.4.595DOI Listing

Publication Analysis

Top Keywords

perfusion parameters
28
intraoperator variability
16
dynamic contrast-enhanced
12
arterial input
12
variability values
12
perfusion
8
quantitative perfusion
8
parameters
8
contrast-enhanced sonography
8
deconvolution method
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!