Background: This study addresses the efficacy of an automated decontamination protocol using the germicide 'tetra acetyl ethylene diamine (TAED) perborate' (Farmec SpA, Italy). The germicide TAED perborate protocol is used in the Castellini Dental Units fitted with an Autosteril unit (an automated device that can cycle 0.26% TAED perborate solution and sterile water for cleaning the water system between patients and overnight). Prior to testing the Autosteril and the 0.26% TAED perborate protocol on the Logos Jr Dental Unit (Castellini SpA, Italy), TAED perborate was used on a dental unit water system simulation device.

Methods: A dental unit water system simulation device equipped with four dental unit water systems and with naturally grown and mature biofilm contamination was used in this study (three treatment units and one control). One treatment group used a simulated 5 minutes contact with TAED perborate and sterile water for irrigation; the second used a simulated 5 minutes contact with TAED perborate and 2 ppm ClO2 for irrigation; the third used a simulated 5 minutes contact with TAED perborate and municipal water for irrigation. The control group used municipal water for irrigation with no cleaning/disinfection protocols. This protocol was repeated for 30 cycles. Laser scanning confocal microscopy (LSCM) was used to study the effects on natural and mature biofilms, and R2A agar used to quantify heterotrophic plate counts in the effluent irrigant. Antimicrobial efficacy was evaluated by challenging TAED perborate with microbes and spores (M. smegmatis and B. subtilis). Deleterious effects of the germicide were evaluated on metal and nonmetal parts of dental unit water systems. Heterotrophic plate counts using R2A agar and LSCM of the lines were conducted to assess biofilm and microbial control.

Results: Baseline water samples showed mean contamination >5.6 log10 cfu/ml. After initial cleaning, all three groups maintained mean contamination levels of less than 1.1 (SD <0.3) log10 cfu/ml. LSCM of baseline samples was positive for live biofilm in all groups. At the end of the study, viable biofilm was only present in the control. In the microbial challenge test, all vegetative organisms were killed within 30 seconds of contact, while spores were killed within 5 minutes. Corrosion was seen in metals used in US-manufactured dental unit materials, while not observed in those used in the Castellini Logos Jr dental unit.

Conclusion: In this study, the TAED perborate protocol was effective in biofilm control and control of dental treatment water contamination. Use of sterile water or 2 ppm ClO2 along with TAED treatment also controlled planktonic contamination effectively.

Clinical Significance: Environmental biofilms contaminate dental unit water systems over time and affect the quality of dental treatment water. Contaminants include environmental biofilms, microbes, including gram-negative rods and endotoxins in high doses that are not of acceptable quality for treating patients. There are many germicidal protocols for treating this contamination and one such is the prescribed use of TAED perborate used in conjunction with sterile water for irrigation in the autosteril device, an integral component of the Castellini dental units for between patient decontamination of dental unit water systems. This study was conducted on an automated simulation dental unit water system to test the TAED perborate protocol's efficacy on naturally grown, mature environmental biofilms, it's efficacy on microbes and spores and it's effects on materials used in dental unit water systems. This translational research addresses both microbial control and material effects of TAED perborate in studying efficacy and possible deleterious effects and simulated use in dentistry. Currently, this antimicrobial use protocol is followed worldwide in the Castellini dental units that are used in day-to-day dental patient care.

Download full-text PDF

Source

Publication Analysis

Top Keywords

taed perborate
32
dental unit
24
unit water
20
water system
12
simulated minutes
12
minutes contact
12
contact taed
12
water irrigation
12
water
11
taed
9

Similar Publications

Background. This study evaluated the antimicrobial activity of Tetraacetylethylenediamine-sodium perborate (TAED-SP) in comparison to 2.5% and 5% sodium hypochlorite (NaOCl) against Enterococcus faecalis.

View Article and Find Full Text PDF

Background: This study addresses the efficacy of an automated decontamination protocol using the germicide 'tetra acetyl ethylene diamine (TAED) perborate' (Farmec SpA, Italy). The germicide TAED perborate protocol is used in the Castellini Dental Units fitted with an Autosteril unit (an automated device that can cycle 0.26% TAED perborate solution and sterile water for cleaning the water system between patients and overnight).

View Article and Find Full Text PDF

Introduction: Tetraacetylethylenediamine in association with sodium perborate (TAED+P) can be suggested for its use as an endodontic disinfectant because of its antimicrobial activity against different bacterial species when used at low concentrations. The purpose of this study was to measure the cytotoxicity of TAED+P on L929 fibroblasts and to compare it with that of sodium hypochlorite (NaOCl).

Methods: L929 fibroblasts were grown in Dulbecco Modified Eagle Medium containing 10% fetal calf serum (FCS) at 37 degrees C and 5% CO(2).

View Article and Find Full Text PDF

The behavior of silver nanotextiles during washing.

Environ Sci Technol

November 2009

Technology and Society Laboratory, and Laboratory of Advanced Fibers, Empa - Swiss Federal Laboratories for Materials Testing and Research, CH 9014 St. Gallen, Switzerland.

The widespread use of silver nanoparticles (Ag-NPs) in commercial products, especially textiles, will likely result in an unknown spread of Ag into the environment. The quantification and characterization of the Ag released from nano-Ag-products is an important parameter needed to predict the effect of Ag-NPs on the environment. The aim of this study was to determine the amount and the form of Ag released during washing from nine fabrics with different ways of silver incorporation into or onto the fibers.

View Article and Find Full Text PDF

Oxygen release of tetra acetyl ethylene diamine (TAED) and sodium perborate combination.

Int Endod J

July 2008

Department of Restorative Dentistry & Endodontics, Ege University School of Dentistry, Bornova, Izmir, Turkey.

Aim: To evaluate the effect of tetra acetyl ethylene diamine (TAED) on oxygen release from sodium perborate and to compare it with sodium perborate-hydrogen peroxide (H(2)O(2)) and sodium perborate-distilled water mixtures.

Methodology: Six groups were evaluated: control groups (groups I-III), sodium perborate was mixed with distilled water or 3% or 30% H(2)O(2); experimental groups (groups IV-VI), sodium perborate was mixed with TAED in different ratios and then distilled water was added to these mixtures. The amount of oxygen released from the samples was measured with a digital oxygen meter at room temperature (25 degrees C) and body temperature (37 degrees C) after 1 min, 1-6 h and 12 h and on each day up to 1 week.