KERMA ratios in pediatric CT dosimetry.

Pediatr Radiol

Department of Radiology and Radiological Science, Medical University of South Carolina (MUSC), Charleston, SC, USA.

Published: May 2012

Background: Patient organ doses may be estimated from CTDI values. More accurate estimates may be obtained by measuring KERMA (Kinetic Energy Released in Matter) in anthropomorphic phantoms and referencing these values to free-in-air X-ray intensity.

Objective: To measure KERMA ratios (R(K)) in pediatric phantoms at CT.

Materials & Methods: CT scans produce an air KERMA K in a phantom and an air KERMA K(CT) at isocenter. KERMA ratios (R(K)) are defined as (K/K(CT)), measured using TLD chips in phantoms representing newborns to 10-year-olds.

Results: R(K) in the newborn is approximately constant. For the other phantoms, there is a peak R(K) value in the neck. The median R(K) values for the GE scanner at 120 kV were 0.92, 0.83, 0.77 and 0.76 for newborns, 1-year-olds, 5-year-olds and 10-year-olds, respectively. Organ R(K) values were 0.91 ± 0.04, 0.84 ± 0.07, 0.74 ± 0.09 and 0.72 ± 0.10 in newborns, 1-year-olds, 5-year-olds and 10-year-olds, respectively. At 120 kV, a Siemens Sensation 16 scanner had R(K) values 5% higher than those of the GE LightSpeed Ultra.

Conclusion: KERMA ratios may be combined with air KERMA measurements at the isocenter to estimate organ doses in pediatric CT patients.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00247-011-2336-4DOI Listing

Publication Analysis

Top Keywords

kerma ratios
16
air kerma
12
kerma
8
ratios pediatric
8
organ doses
8
newborns 1-year-olds
8
1-year-olds 5-year-olds
8
5-year-olds 10-year-olds
8
values
5
pediatric dosimetry
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!