Continuous glucose monitoring--a study of the Enlite sensor during hypo- and hyperbaric conditions.

Diabetes Technol Ther

Göteborg Pediatric Growth Research Center, Department of Pediatrics, Institute of Clinical Sciences, Sahlgrenska Academy at University of Gothenburg, Göteborg, Sweden.

Published: June 2012

Background: The performance and accuracy of the Enlite(™) (Medtronic, Inc., Northridge, CA) sensor may be affected by microbubble formation at the electrode surface during hypo- and hyperbaric conditions. The effects of acute pressure changes and of prewetting of sensors were investigated.

Materials And Methods: On Day 1, 24 sensors were inserted on the right side of the abdomen and back in one healthy individual; 12 were prewetted with saline solution, and 12 were inserted dry. On Day 2, this procedure was repeated on the left side. All sensors were attached to an iPro continuous glucose monitoring (CGM) recorder. Hypobaric and hyperbaric tests were conducted in a pressure chamber, with each test lasting 105 min. Plasma glucose values were obtained at 5-min intervals with a HemoCue(®) (Ängelholm, Sweden) model 201 glucose analyzer for comparison with sensor glucose values.

Results: Ninety percent of the CGM systems operated during the tests. The mean absolute relative difference was lower during hyperbaric than hypobaric conditions (6.7% vs. 14.9%, P<0.001). Sensor sensitivity was slightly decreased (P<0.05) during hypobaric but not during hyperbaric conditions. Clarke Error Grid Analysis showed that 100% of the values were found in the A+B region. No differences were found between prewetted and dry sensors.

Conclusions: The Enlite sensor performed adequately during acute pressure changes and was more accurate during hyperbaric than hypobaric conditions. Prewetting the sensors did not improve accuracy. Further studies on type 1 diabetes subjects are needed under various pressure conditions.

Download full-text PDF

Source
http://dx.doi.org/10.1089/dia.2011.0284DOI Listing

Publication Analysis

Top Keywords

continuous glucose
8
hypo- hyperbaric
8
hyperbaric conditions
8
glucose monitoring--a
4
monitoring--a study
4
study enlite
4
enlite sensor
4
sensor hypo-
4
hyperbaric
4
conditions background
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!