[Quality control of dose delivered by in vivo dosimetry: can one tolerance be used for all localizations?].

Cancer Radiother

Service de radiothérapie, hôpital européen Georges-Pompidou, Paris, France.

Published: April 2012

Purpose: In vivo dosimetry measurements are accepted when the difference between measured and calculated dose is under 5%. A statistical analysis has been conducted to determine whether this tolerance matched the clinical practice for the studied localizations: pelvis, thorax, head and neck, breast.

Materials And Methods: The technical characteristics of the detectors were checked before being used in clinical practice. Then an automatic statistical analysis was implemented using the 2450 in vivo dosimetry measurements obtained during 1 year.

Main Results: The global average is 1.10%, the standard deviation 2.46% and the percentage of out of level measurements 4.09%. By distinguishing the localizations, the 5% tolerance appeared to be too narrow for the breast localization.

Discussion/conclusion: Several investigations were initiated to justify the modification of the tolerance for the breast localization. They highlighted an underestimation of the calculated dose when high beam angles are set: a new correction factor was defined to take account this error. A specific tolerance was also specified for the breast localization.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.canrad.2011.10.010DOI Listing

Publication Analysis

Top Keywords

vivo dosimetry
12
dosimetry measurements
8
calculated dose
8
statistical analysis
8
clinical practice
8
tolerance breast
8
breast localization
8
tolerance
5
[quality control
4
control dose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!