A 12-year-old boy with Marfan's syndrome required a biventricular assist device (VAD) after an aortic root replacement. The patient developed acute respiratory distress syndrome and required escalating ventilator support. We hypothesized that the addition of a membrane oxygenator in series with the assist device would improve gas exchange and allow for a more lung-protective ventilator approach. A membrane oxygenator was placed in series with the right VAD resulting in a blood path of right atrium to VAD to oxygenator to pulmonary artery. Circuit function was gauged by monitoring flow and oxygenator pressures and periodic circuit inspections and oxygenator blood gases. Heparin was titrated to maintain unfractionated antifactor Xa levels of .3-.7 IU/mL and partial thromboplastin time of 60-80 seconds. The initial sweep gas supplying the oxygenator was 5 L/min at an F1O2 of 1.0, which achieved a pH > 7.40 and a PF ratio > 250. The pre- and post-oxygenator pressures were 55-60 mmHg and 45-50 mmHg, respectively, and the measured flow at the oxygenator outlet was 2.0-2.2 L/min. The patient was changed from high-frequency oscillatory ventilation to pressure-controlled synchronized intermittent ventilation with pH maintained at 7.35-7.40 and PF ratio > 250. Paralytics were discontinued and the patient's neurologic condition was deemed intact. The patient hemorrhaged after a sternal closure and required transfusions and antifibrinolytics that led to thrombus in the membrane and membrane circuitry, which were replaced without incident. The patient's respiratory status remained stable; however, his overall condition worsened as a result of additional organ dysfunction and septicemia, and he did not survive. The addition of a membrane oxygenator to a VAD is feasible and supplements gas exchange permitting the use of more lung protective ventilation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4557432 | PMC |
Artif Organs
January 2025
Division of Life Science and Medicine, School of Biomedical Engineering (Suzhou), University of Science and Technology of China, Hefei, China.
Background: Membrane oxygenators facilitate extracorporeal gas exchange, necessitating the monitoring of blood gas. Recent advances in normothermic machine perfusion (NMP) for ex vivo liver offer solutions to the shortage of donor liver. However, maintaining physiological blood gas levels during prolonged NMP is complex and costly.
View Article and Find Full Text PDFBMC Anesthesiol
January 2025
Department of Nephrology, University of Duisburg-Essen, University Hospital Essen, Essen, Germany.
Introduction: Acute kidney injury (AKI) is a common complication of acute respiratory distress syndrome (ARDS) and multiple organ dysfunction syndrome (MODS) in patients receiving extracorporeal membrane oxygenation (ECMO) support, leading to requirement of continuous renal replacement therapy (CRRT) in 70% of ECMO patients. Parallel arrangement of CRRT and ECMO circuits is common in adult patients. However, CRRT may also be integrated directly into the ECMO circuit.
View Article and Find Full Text PDFMacromol Biosci
January 2025
Institute for Bioengineering of Catalonia (IBEC), The Barcelona Institute of Science and Technology (BIST), Carrer de Baldiri Reixac, 10, 12, Barcelona, 08028, Spain.
Blood-contacting medical devices, especially extracorporeal membrane oxygenators (ECMOs), are highly susceptible to surface-induced coagulation because of their extensive surface area. This can compromise device functionality and lead to life-threatening complications. High doses of anticoagulants, combined with anti-thrombogenic surface coatings, are typically employed to mitigate this risk, but such treatment can lead to hemorrhagic complications.
View Article and Find Full Text PDFJ Artif Organs
January 2025
Department for Cardiac, Thoracic and Cardiovascular Surgery, University Hospital Regensburg, Regensburg, Germany.
Neutrophil extracellular traps (NETs) were detected in blood samples and in cellular deposits of oxygenator membranes during extracorporeal membrane oxygenation (ECMO) therapy and may be responsible for thrombogenesis. The aim was to evaluate the effect of the base material of gas fiber (GF, polymethylpentene) and heat exchange (HE) membranes and different antithrombogenic coatings on isolated granulocytes from healthy volunteers under static culture conditions. Contact of granulocytes with membranes from different ECMO oxygenators (with different surface coatings) and uncoated-GFs allowed detection of adherent cells and NETotic nuclear structures (normal, swollen, ruptured) using nuclear staining.
View Article and Find Full Text PDFJ Clin Med
December 2024
Department of Surgery IV, George Emil Palade University of Medicine, Pharmacy, Science and Technology of Targu Mures, 38 Gheorghe Marinescu Street, 540139 Targu Mures, Romania.
In recent decades, the advantages of minimizing surgical trauma have led to the development of minimally invasive surgical procedures. While the benefits often outweigh the risks, several challenges are encountered that are not present in conventional surgical approaches. Unilateral pulmonary edema (UPE) after mitral interventions performed through a right-sided approach is a rare but potentially life-threatening event.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!