Obesity increases precision errors in dual-energy X-ray absorptiometry measurements.

J Clin Densitom

Department of Medical Imaging, College of Engineering, Mathematics and Physical Sciences, University of Exeter, Exeter, UK.

Published: September 2012

The precision errors of dual-energy X-ray absorptiometry (DXA) measurements are important for monitoring osteoporosis. This study investigated the effect of body mass index (BMI) on precision errors for lumbar spine (LS), femoral neck (NOF), total hip (TH), and total body (TB) bone mineral density using the GE Lunar Prodigy. One hundred two women with BMIs ranging from 18.5 to 45.9 kg/m(2) were recruited. Participants had duplicate DXA scans of the LS, left hip, and TB with repositioning between scans. Participants were divided into 3 groups based on their BMI and the percentage coefficient of variation (%CV) calculated for each group. The %CVs for the normal (<25 kg/m(2)) (n=48), overweight (25-30 kg/m(2)) (n=26), and obese (>30 kg/m(2)) (n=28) BMI groups, respectively, were LS BMD: 0.99%, 1.30%, and 1.68%; NOF BMD: 1.32%, 1.37%, and 2.00%; TH BMD: 0.85%, 0.88%, and 1.06%; TB BMD: 0.66%, 0.73%, and 0.91%. Statistically significant differences in precision error between the normal and obese groups were found for LS (p=0.0006), NOF (p=0.005), and TB BMD (p=0.025). These results suggest that serial measurements in obese subjects should be treated with caution because the least significant change may be larger than anticipated.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jocd.2012.01.002DOI Listing

Publication Analysis

Top Keywords

precision errors
12
errors dual-energy
8
dual-energy x-ray
8
x-ray absorptiometry
8
bmd
5
obesity increases
4
precision
4
increases precision
4
absorptiometry measurements
4
measurements precision
4

Similar Publications

This paper presents a coaxial integrated macro-micro composite actuator. The macro-actuator of the macro-micro composite actuator is similar to a moving coil type voice coil motor, and a giant magnetostrictive actuator is installed coaxially inside it as a micro-actuator. In this work, kinetic models are established for both the macro-actuator and micro-actuator, and based on the models, an automatic disturbance rejection controller is adopted to control the macro-actuator, and a fuzzy sliding mode controller is adopted for the micro-actuator.

View Article and Find Full Text PDF

Ultra-precise ruler for ammonia nitrogen quantification in electrochemical synthesis experiments.

Anal Methods

January 2025

Key Laboratory of Synthetic and Biological Colloids, Ministry of Education, School of Chemical and Material Engineering, Jiangnan University, Wuxi, Jiangsu 214122, P. R. China.

The field of electrochemical ammonia synthesis has made rapid advancements, attracting a large number of scientists to contribute to this area of research. Accurate detection of ammonia is crucial in this process for evaluating the efficiency and selectivity of electrocatalysts. In this study, we systematically investigate the indophenol blue method for ammonia detection, examining the effects of key factors such as solution pH, nitrate concentration, and metal ion concentration on measurement accuracy.

View Article and Find Full Text PDF

Introduction: Sacroiliac luxation is a common traumatic feline injury, with the small size of the sacral body being a challenge for surgical stabilization. This study compared an innovative computer-guided drilling method with the conventional fluoroscopy-controlled freehand technique. Neuronavigation, using CT-based planning and real-time tracking, was evaluated against the freehand method for accuracy and time efficiency.

View Article and Find Full Text PDF

Background: Body composition scores allow for quantifying the volume and physical properties of specific tissues. However, their manual calculation is time-consuming and prone to human error. This study aims to develop and validate CompositIA, an automated, open-source pipeline for quantifying body composition scores from thoraco-abdominal computed tomography (CT) scans.

View Article and Find Full Text PDF

Non-optically active water quality parameters (NAWQPs) are essential for surface water quality assessments, although automated monitoring methods are time-consuming, include labor-intensive chemical pretreatment, and pose challenges for high spatiotemporal resolution monitoring. Advancements in spectroscopic techniques and machine learning may address these issues. We integrated ultraviolet-visible-near infrared absorption spectroscopy with physical-chemical measurements to predict total nitrogen (TN), dissolved oxygen (DO), and total phosphorus (TP) in the Yangtze River Basin, China.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!