  • Titanium implants face issues like low wear resistance and the release of titanium particles into surrounding tissue, but these can be improved with surface engineering methods like nitriding.
  • The study focuses on combining nitriding and oxidizing treatments at 680 degrees Celsius to create a diffusive protective layer of TiO + TiN + Ti2N + alphaTi(N), enhancing wear and corrosion resistance.
  • Results show that this new layered surface promotes better biocompatibility, with increased adhesion and proliferation of osteoblasts, making it a promising option for long-term bone implants.

Article Abstract

The disadvantages of titanium implants are their low wear resistance and the release of titanium elements into surrounding tissue. These can be eliminated by modifying the surface by surface engineering methods, among them nitriding under glow discharge conditions which allow to produce diffusive surface layers. Their combining with an oxide layer might be valuable for biological events occurring at the bone implant interface. The aim of this study was to enhance the titanium biomaterial performance via combining nitriding and oxidizing treatments in one process under glow discharge conditions. The oxynitrided surface layers were produced at 680 degrees C. The obtained layer was TiO + TiN + Ti2N + alphaTi(N) type and about 4-microm thick and was of diffusive character. This layer significantly increased wear resistance and slightly corrosion resistance compared to that of the reference titanium alloy. The produced titanium oxide was about 400-nm thick and built from fine crystallites. This oxide exhibits bioactivity in SBF (simulated body fluid). Osteoblasts of Saos-2 line incubated on this surface exhibited good adhesion and proliferation and ALP release comparable with cells cultured on the reference titanium alloy and TiN + Ti2N + alphaTi(N) surface layers. A quantitative analysis of blood platelets adhering to this layer revealed their highest amount in comparison to that on both the nitrided surface layer and titanium alloy. The presented study provided a simple and reproducible method of combining oxidizing and nitriding under glow discharge in one process. Experimental data in vitro suggests that titanium alloy oxynitriding under low temperatures at glow discharge conditions improves titanium alloy properties and biocompatibility and tissue healing. Therefore, the layer of TiO + TiN +Ti2N + alphaTi(N) type could be valuable for long-term bone implants.

Download full-text PDF

Source
http://dx.doi.org/10.1166/jnn.2011.3474DOI Listing

Publication Analysis

Top Keywords

glow discharge
20
titanium alloy
20
discharge conditions
16
surface layers
12
titanium
10
surface
8
nitrided surface
8
surface layer
8
wear resistance
8
nitriding glow
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!