Background: Spinal hardware has been adapted for fixation in the setting of anterior pelvic injury. This anterior subcutaneous pelvic fixator consists of pedicle screws placed in the supraacetabular region connected by a contoured connecting rod placed subcutaneously and above the abdominal muscle fascia.

Questions/purposes: We examined the placement of the components for anterior subcutaneous pelvic fixator relative to key vascular, urologic, bony, and surface structures.

Methods: We measured the CT scans of 13 patients after placement of the pelvic fixator to determine the shortest distances between the fixator components and important anatomic structures: the femoral vascular bundle, the urinary bladder, the cranial margin of the hip, the screw insertion point on the bony pelvis, the relationship between the pedicle screw and the corridor of bone in which it resided, and the position relative to the skin.

Results: The average distance from the vascular bundle to the pedicle screw was 4.1 cm and 2.2 cm to the connecting rod. The average distance from the connecting rod to the anterior edge of the bladder was 2.6 cm. The average distance from the screw insertion point to the hip was 2.4 cm; none penetrated the hip. The average screw was in bone for 5.9 cm. The pedicle screws were on average 2.1 cm under the skin. The average distance from the anterior skin to the connecting rod was 2.7 cm.

Conclusions: Components of this anterior pelvic fixator are close to important anatomic structures. Careful adherence to the surgical technique should minimize potential risk.

Level Of Evidence: Level IV, retrospective study. See Guidelines for Authors for a complete description of levels of evidence.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3392368PMC
http://dx.doi.org/10.1007/s11999-012-2287-6DOI Listing

Publication Analysis

Top Keywords

pelvic fixator
20
connecting rod
16
average distance
16
anterior pelvic
12
anterior subcutaneous
8
subcutaneous pelvic
8
pedicle screws
8
components anterior
8
anatomic structures
8
vascular bundle
8

Similar Publications

Abdominal wall repair in adults with bladder exstrophy is challenging. We present a case of a 46-year-old woman with bladder exstrophy presenting with a large midline incisional hernia associated with a 13-cm hypoplasia of both pubic rami that precluded fixation of any abdominal mesh. A two-stage approach was adopted.

View Article and Find Full Text PDF

Case: A 35-year-old man with morbid obesity sustained an ultra-low velocity (ULV) rotational knee dislocation secondary to a fall from standing. The patient was successfully treated using a subcutaneous knee-spanning internal fixator, the "INFIX" technique, which has previously been described for pelvic ring injuries.

Conclusion: This novel technique maintained the stable reduction of an ULV knee dislocation in a patient with morbid obesity until adequate healing was achieved.

View Article and Find Full Text PDF

Total laparoscopic closed reduction and internal fixation for AO/OTA B2.1 pelvic fracture: A case report and literature review.

Chin J Traumatol

December 2024

Division of Trauma and War Injury, Daping Hospital, Army Medical University of PLA, Chongqing, 400042, China. Electronic address:

Endoscopic techniques have been widely used in orthopedic surgery, such as arthroscopy and transforaminal endoscopy, but the application in fracture is rarely reported. We reported a case of a 69-year-old male with pelvic fracture (AO/OTA type B2.1) who underwent successful laparoscopy-assisted pubic ramus plate fixation without auxiliary incision.

View Article and Find Full Text PDF

Background: Hemodynamically unstable pelvic ring fractures from high-energy trauma are critical injuries in trauma care, requiring urgent intervention and precise diagnostics. With ongoing advancements in trauma management, treatment strategies have evolved, with some techniques becoming obsolete as new ones emerge. This study aimed to evaluate changes and trends in treatment algorithms for these injuries over approximately 40 years.

View Article and Find Full Text PDF

Background: Pelvic reconstruction after type I + II (or type I + II + III) internal hemipelvectomy with extensive ilium removal is a great challenge. In an attempt to anatomically reconstruct the hip rotation center (HRC) and achieve a low mechanical failure rate, a custom-made, 3D-printed prosthesis with a porous articular interface was developed. The aim of this study was to investigate the clinical outcomes of patients treated with this prosthesis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!