Background: In 18F-fluorodeoxyglucose PET/CT, respiratory motion induces bias in image interpretations (i.e. organ misregistration or lesion omission/underestimation).

Objective: The present study applied our custom gating method (referred to as 'CT-based gated PET' and that has already been validated by our research group) to PET imaging of the liver and compared its per lesion sensitivity with that of standard clinical (i.e. ungated) PET.

Methods: A total of 13 patients scheduled for liver surgery were referred to our department for PET/CT imaging. Each patient underwent both ungated and CT-based gated PET imaging protocols. Two independent, blinded observers interpreted the two sets of PET images and reached a consensus when necessary. Image interpretations were combined with histological analysis and/or intraoperative ultrasound examination to compute each method's per lesion sensitivity and true positive fraction. Analyses were also performed by considering lesions according to their size (longest axis over 10 mm or over 15 mm).

Results: Forty-eight lesions were confirmed by pathology reports. When considering all the uptakes, the ungated and the CT-based gated PET methods had sensitivities of 54.2 and 64.6% (P=0.025) and true positive fractions of 83.9 and 86.1%, respectively. When considering uptakes greater than 10 mm in size, ungated and CT-based gated PET had sensitivities of 74.3 and 88.6% (P=0.025), respectively. For lesions greater than 15 mm in size, the corresponding sensitivities were 85.2 and 100% (P=0.046).

Conclusion: CT-based gated PET yielded a higher lesion-based sensitivity than routine, ungated PET did. Hence, this method improves the detection of intrahepatic colorectal metastases--especially for lesions that are close in size to the detection limit of the PET gantry.

Download full-text PDF

Source
http://dx.doi.org/10.1097/MNM.0b013e328351fce8DOI Listing

Publication Analysis

Top Keywords

ct-based gated
16
gated pet
16
ungated ct-based
12
intrahepatic colorectal
8
image interpretations
8
pet
8
pet imaging
8
lesion sensitivity
8
true positive
8
considering uptakes
8

Similar Publications

GFR Estimation and Correlation for Oncology Patients by Two Methods, Gates Method and Dual Time Point Plasma Sampling Method.

Indian J Nucl Med

November 2024

Department of Nuclear Medicine and Molecular Imaging, Homi Bhabha Cancer Hospital and Mahamana Pandit Madan Mohan Malaviya Cancer Centre, Tata Memorial Centre, Homi Bhabha National Institute (HBNI), Varanasi, India.

Background: With the increasing number of oncology cases and a parallel surge in chemotherapeutic drugs for treatment, the treating physicians conducts nephrotoxicity evaluation to provide a personalized dosing strategy. Of the various tests available, glomerular filtration rate (GFR) under gamma camera with help of Gates method has gained importance, being a good index of overall kidney functions. In addition to this, there has been an alternate and old method for GFR estimation: plasma sampling.

View Article and Find Full Text PDF

Whole-body PET imaging is often hindered by respiratory motion during acquisition, causing significant degradation in the quality of reconstructed activity images. An additional challenge in PET/CT imaging arises from the respiratory phase mismatch between CT-based attenuation correction and PET acquisition, leading to attenuation artifacts. To address these issues, we propose two new, purely data-driven methods for the joint estimation of activity, attenuation, and motion in respiratory self-gated TOF PET.

View Article and Find Full Text PDF

Background: Regional myocardial work (MW) is not measured in the right ventricle (RV) due to a lack of high spatial resolution regional strain (RS) estimates throughout the ventricle. We present a cineCT-based approach to evaluate regional RV performance and demonstrate its ability to phenotype three complex populations: end-stage LV failure (HF), chronic thromboembolic pulmonary hypertension (CTEPH), and repaired tetralogy of Fallot (rTOF).

Methods: 49 patients (19 HF, 11 CTEPH, 19 rTOF) underwent cineCT and right heart catheterization (RHC).

View Article and Find Full Text PDF

The use of bioprostheses is increasing in younger patients, but it is associated with the risk of later valve deterioration, especially in the mitral position. A new bioprosthesis for mitral valve replacement offers possible longer-term durability and improved hemodynamics. : Here, we report the implantation of the novel Edwards MITRIS RESILIA mitral valve (Edwards Lifesciences Inc.

View Article and Find Full Text PDF

Introduction: Whole-body plethysmography is the preferred method for measuring the static lung volumes: total lung capacity (TLC), functional residual capacity (FRC) and residual volume (RV), as it also incorporates trapped gas - a common finding in chronic obstructive pulmonary disease (COPD). Quantitative computed tomography (CT) is a promising alternative to plethysmography, which can be challenging to perform for patients with severely impaired lung function. The present systematic review explores the agreement between lung volumes measured by plethysmography and CT, as well as the attempts being made to optimize alignment between these two methods.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!