Object: Vagus nerve stimulation (VNS) has demonstrated benefit in patients with medically intractable partial epilepsy. As in other therapies with mechanical devices, hardware failure occurs, most notably within the VNS lead, requiring replacement. However, the spiral-designed lead electrodes wrapped around the vagus nerve are often encased in dense scar tissue hampering dissection and removal. The objective in this study was to characterize VNS lead failure and lead revision surgery and to examine VNS efficacy after placement of a new electrode on the previously used segment of vagus nerve.

Methods: The authors reviewed all VNS lead revisions performed between October 2001 and August 2011 at the University of Iowa Hospitals and Clinics. Twenty-four patients underwent 25 lead revisions. In all cases, the helical electrodes were removed, and a new lead was placed on the previously used segment of vagus nerve. All inpatient and outpatient records of the 25 lead revisions were retrospectively reviewed.

Results: Four cases were second lead revisions, and 21 cases were first lead revisions. The average time to any revision was 5 years (range 1.8-11.1 years), with essentially no difference between a first and second lead revision. The most common reason for a revision was intrinsic lead failure resulting in high impedance (64%), and the most common symptom was increased seizure frequency (72%). The average duration of surgery for the initial implantation in the 15 patients whose VNS system was initially implanted at the authors' institution was much shorter (94 minutes) than the average duration of lead revision surgery (173 minutes). However, there was a significant trend toward shorter surgical times as more revision surgeries were performed. Sixteen of the 25 cases of lead revision were followed up for more than 3 months. In 15 of these 16 cases, the revision was as effective as the previous VNS lead. In most of these cases, both the severity and frequency of seizures were decreased to levels similar to those following the previous implantation procedure. Only 1 complication occurred, and there were no postoperative infections.

Conclusions: Lead revision surgery involving the placement of a new electrode at the previously used segment of vagus nerve is effective at decreasing the seizure burden to an extent similar to that obtained following the initial VNS implantation. Even with multiple lead revisions, patients can obtain VNS efficacy similar to that following the initial lead implantation. There is a learning curve with revision surgery, and overall the duration of surgery is longer than for the initial implantation. Note, however, that complications and infection are rare.

Download full-text PDF

Source
http://dx.doi.org/10.3171/2012.1.FOCUS11333DOI Listing

Publication Analysis

Top Keywords

lead revision
24
lead revisions
24
vagus nerve
20
lead
19
vns lead
16
revision surgery
16
segment vagus
12
revision
11
vns
9
nerve stimulation
8

Similar Publications

Background: The differentiation between association and causation is a significant challenge in medical research, often further complicated by cognitive biases that erroneously interpret coincidental observational data as indicative of causality. Such misinterpretations can lead to misguided clinical guidelines and healthcare practice, potentially endangering patient safety and leading to inefficient use of resources.

Methods: We conducted an extensive search of PubMed, Cochrane, and Embase databases up to March 2024, identifying circumstances where associations from observational studies were incorrectly deemed causal.

View Article and Find Full Text PDF

Interprofessional Collaboration and Patient/Family Engagement on Rounds in a Comprehensive Stroke Center: A Mixed-Methods Study.

Qual Manag Health Care

December 2024

Author Affiliations: Armstrong Institute for Patient Safety and Quality, Johns Hopkins University School of Medicine, Baltimore, Maryland (Dr Xie); Departments of Anesthesiology and Critical Care Medicine (Dr Xie) and Neurology (Drs Bahouth, Salas, and Zink), Johns Hopkins University School of Medicine, Baltimore, Maryland; Department of Nursing, The Johns Hopkins Hospital, Baltimore, Maryland (Drs Barany, Watson, Zink, and Hairston and Ms Shakes); Johns Hopkins University School of Nursing (Drs Tanner, Hanson, Hansen, McDonald, and Hairston), Baltimore, Maryland; and Department of Biobehavioral Nursing and Health Informatics, School of Nursing, University of Washington, Seattle, Washington (Dr Abu-Rish Blakeney).

Background And Objectives: Daily rounds provide an opportunity for interprofessional collaboration and patient/family engagement, which are critical to stroke care. As part of a quality improvement program, we conducted a baseline assessment to examine interprofessional collaboration and patient/family engagement during the current rounding process in a 12-bed comprehensive stroke center. Findings from the baseline assessment will be used to inform the development, implementation, and evaluation of a new rounding model.

View Article and Find Full Text PDF

Background: In rare circumstances, an implanted deep brain stimulation device will develop impedance issues across its contacts. Even more rare is the resultant inability to program a patient effectively, or that the patients' prior programming settings become unusable.

Objective: In this study we investigate this occurrence across the device manufacturers implanted, and whether this could be resolved.

View Article and Find Full Text PDF

Background: Unicompartmental knee arthroplasty (UKA) procedures have become much more common in the United States in recent years, with >40,000 UKAs performed annually. However, it is estimated that 10% to 40% of UKAs fail and thus require conversion to total knee arthroplasty (TKA). In the field of total joint arthroplasty, robotic-assisted surgeries have demonstrated advantages such as better accuracy and precision of implant positioning and improved restoration of a neutral mechanical axis.

View Article and Find Full Text PDF

species have commonly been reported as important plant pathogenic fungi with wide host ranges and geographic distributions. With the increase in the number of cryptic species being described, a comprehensive global taxonomic revision of the genus is required. The present study includes 399 isolates from 32 countries.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!