Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: Patients with major burns accompanied with airway edema need more attention for airway management. Although the Pentax-AWS has an advantage in managing endotracheal intubation more easily, its usefulness cannot be assured if it does not maintain hemodynamic stability in burn patients. The aim of this study was to compare cardiovascular responses and general efficacy of the Pentax-AWS and Macintosh laryngoscopes in burn patients.
Methods: American Society of Anesthesiologists physical status 2 or 3 adult patients with major burn injury were randomly assigned to group P (AWS, n = 50) or group M (Macintosh, n = 50). Fifty-nine patients assigned to the Macintosh group and no patient to AWS group were excluded because of failure to first intubation. Hemodynamic data at baseline, just before and after intubation as well as 3, 5 and 10 minutes after intubation and grade of sore throat were recorded in two groups. Intubation time, success rate of intubation, number of intubation attempts and the percentage of glottic opening (POGO) scores were also observed and analyzed in all patients.
Results: A statistically significant increase in heart rate just after laryngoscopy was seen in group M. The success rate of the first attempt in group P (100%) was greater than with the group M (46%). POGO scores were higher in group P (97 ± 4%) than in group M (48 ± 29%)
Conclusions: The use of Pentax-AWS offers a reduced degree of heart rate elevation compared with the Macintosh laryngoscope and better intubation environments in major burn patients.