High spatial resolution Kelvin probe force microscopy with coaxial probes.

Nanotechnology

School of Engineering and Applied Sciences, Harvard University, Cambridge, MA 02138, USA.

Published: March 2012

Kelvin probe force microscopy (KPFM) is a widely used technique to measure the local contact potential difference (CPD) between an AFM probe and the sample surface via the electrostatic force. The spatial resolution of KPFM is intrinsically limited by the long range of the electrostatic interaction, which includes contributions from the macroscopic cantilever and the conical tip. Here, we present coaxial AFM probes in which the cantilever and cone are shielded by a conducting shell, confining the tip-sample electrostatic interaction to a small region near the end of the tip. We have developed a technique to measure the true CPD despite the presence of the shell electrode. We find that the behavior of these probes agrees with an electrostatic model of the force, and we observe a factor of five improvement in spatial resolution relative to unshielded probes. Our discussion centers on KPFM, but the field confinement offered by these probes may improve any variant of electrostatic force microscopy.

Download full-text PDF

Source
http://dx.doi.org/10.1088/0957-4484/23/11/115703DOI Listing

Publication Analysis

Top Keywords

spatial resolution
12
force microscopy
12
kelvin probe
8
probe force
8
technique measure
8
electrostatic force
8
electrostatic interaction
8
force
5
probes
5
electrostatic
5

Similar Publications

A proof-of-concept study for precise mapping of pigmented basal cell carcinoma in asian skin using multispectral optoacoustic tomography imaging with level set segmentation.

Eur J Nucl Med Mol Imaging

January 2025

A*STAR Skin Research Labs (A*SRL), Agency for Science, Technology and Research (A*STAR), 31 Biopolis Way, #07-01, Nanos, Singapore, 138669, Republic of Singapore.

Purpose: Basal Cell Carcinoma (BCC), the most common subtype of non-melanoma skin cancers (NMSC), is prevalent worldwide and poses significant challenges due to their increasing incidence and complex treatment considerations. Existing clinical approaches, such as Mohs micrographic surgery, are time-consuming and labour-intensive, requiring meticulous layer-by-layer excision and examination, which can significantly extend the duration of the procedure. Current optical imaging solutions also lack the necessary spatial resolution, penetration depth, and contrast for effective clinical use.

View Article and Find Full Text PDF

This experimental phantom study investigates current standard of care protocols in cone beam computed tomography (CBCT), energy-integrating-detector (EID) CT, and photon-counting-detector (PCD) CT regarding their potential in delineation of dental root canals. Artificial accessory canals (diameters: 1000, 600, 400, 300 and 200 μm) were drilled into three bovine teeth mounted on a bovine rib as a jaw substitute. The phantom was scanned in two dental CBCTs, two EID-CTs and a PCD-CT using standard clinical protocols.

View Article and Find Full Text PDF

High costs and project-based (short-term) financing mean that coastal engineering projects are often undertaken in the absence of appropriate post-construction monitoring programmes. Consequently, the performance of shoreline-stabilizing structures or beach nourishments cannot be properly quantified. Given the high value of beaches and the increase in erosion problems and coastal engineering responses, managers require as much accurate data as possible to support efficient decision-making.

View Article and Find Full Text PDF

The need for effective early detection and optimal therapy monitoring of cardiovascular diseases as the leading cause of death has led to an adaptation of the guidelines with a focus on cardiac computed tomography (CCTA) in patients with a low to intermediate risk of coronary heart disease (CHD). In particular, the introduction of photon-counting computed tomography (PCCT) in CT diagnostics promises significant advances through higher temporal and spatial resolution, and also enables advanced texture analysis, known as radiomics analysis. Originally developed in oncological imaging, radiomics analysis is increasingly being used in cardiac imaging and research.

View Article and Find Full Text PDF

Owing to its high spatial resolution and its high sensitivity to chemical element detection, transmission electron microscopy (TEM) technique enables to address high-level materials characterization of advanced technologies in the microelectronics field. TEM instruments fitted with various techniques are well-suited for assessing the local structural and chemical order of specific details. Among these techniques, 4D-STEM is suitable to estimate the strain distribution of a large field of view.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!