Purpose: Among other corneal biomechanical properties, Goldmann applanation tonometry (GAT) has been shown to depend on corneal edema. New tonometry devices have been designed, such as the Tono-Pen XL, iCare, and ocular response analyzer (ORA), to measure the intraocular pressure (IOP) accurately. This study aims to investigate the influence of corneal edema on the accuracy of these IOP-measuring devices in an in vitro model.
Methods: A model of an artificial anterior chamber was developed using a guided trephination system. Eight donor corneas not suitable for keratoplasty were clamped into this artificial anterior chamber. All corneas showed signs of stromal edema. Intracameral pressure (ICP) was adjusted manometrically to 10, 20, 30, 40, and 50 mm Hg. The central corneal thickness (CCT) was determined by ultrasonic pachymetry. For each manometrically defined ICP, tonometry was performed using the iCare, Tono-Pen XL, GAT, and ORA.
Results: The mean CCT increased from 616.1±29.6 µm to 626.9±36.1 µm. At 10 mm Hg, GAT yielded a higher ICP than those manometrically adjusted (10.4±3.3 mm Hg); at all other ICP levels, GAT yielded lower ICP levels than those adjusted. The Tono-Pen XL and iCare showed the greatest difference at 10 mm Hg, with the Tono-Pen XL yielding a value of 14.0±4.0 mm Hg and the iCare yielding a value of 12.5±2.6 mm Hg. All other results of the 2 devices fell within a range of ±2 mm Hg from the adjusted ICP. The ORA provided accurate results only at "physiological" ICP levels with a maximum difference of 2.6 mm Hg at 30 mm Hg. At higher ICP levels, corneal hysteresis decreased significantly with increasing ICP. None of the measurement devices revealed a statistically relevant dependence on CCT in this experimental setting.
Conclusions: The Tono-Pen XL and the iCare yielded the most accurate ICP values across all the adjusted ICP values. This may be because of their relatively small contact area with the cornea and, consequently, greater independence from corneal biomechanical properties. The ORA yielded accurate measurement results only at physiological ICP levels. As anticipated, GAT underestimated ICP. The Tono-Pen XL and the iCare should therefore be used to determine IOP in patients suffering from corneal edema, such as bullous keratopathy or Fuchs endothelial dystrophy.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/IJG.0b013e31824cef11 | DOI Listing |
Clin Ophthalmol
November 2023
Department of Ophthalmology, Chang Gung Memorial Hospital, Chiayi, 61363, Taiwan.
Vet Ophthalmol
September 2023
Department of Ophthalmology and Visual Sciences, School of Medicine and Public Health, University of Wisconsin, Madison, Wisconsin, USA.
Objectives: To determine the accuracy, precision, and clinical applicability of the ICare® TONOVET Plus (TVP) in cats.
Animals And Procedures: IOP readings obtained with the TVP were compared to values obtained concurrently with the original TONOVET (TV01) and Tono-Pen Vet™ (TP) in 12 normal cats (24 eyes) and 8 glaucomatous LTBP2-mutant cats (13 eyes) in vivo. Reproducibility of TVP readings was also assessed for three observers in the above cats.
J Clin Med
October 2022
Department of Medicine, Surgery and Dentistry, Scuola Medica Salernitana, University of Salerno, 84131 Salerno, Italy.
Background: Intraocular pressure (IOP) is influenced by body position. The purpose of this study is to compare the IOP measurements obtained with two different devices, to investigate IOP changes in standing, sitting, and supine positions. Methods: In this comparative prospective case series, IOP was measured in sitting, supine, prone, and standing (standing 1) positions and again five minutes after standing (standing 2), utilizing an Icare Pro (ICP) and a Tono-Pen Avia (TPA) in the 64 eyes of 32 healthy subjects.
View Article and Find Full Text PDFVestn Oftalmol
March 2022
Research Institute of Eye Diseases, Moscow, Russia.
Changes in intraocular pressure have great influence on the course of various processes in ocular tissues. Presently, the scientific discussion has returned to the theoretical issues of tonometry and the creation of models that would allow assessing the influence of various ocular parameters on IOP values. Some of the modern tonometers (Icare, Tono-Pen) are convenient and effective when measuring IOP in eyes with damaged, uneven or edematous cornea.
View Article and Find Full Text PDFDiagnostics (Basel)
January 2022
Department of Ophthalmology, University of Medicine and Pharmacy "Carol Davila", 050474 Bucharest, Romania.
Assessing the intraocular pressure is a difficult but crucial task in the follow-up of patients that have undergone penetrating keratoplasty. Early recognition of elevated intraocular pressure and/or glaucoma and establishment of the appropriate treatment is essential to ensure the best possible visual outcome for patients dealing with this feared complication. Although Goldmann applanation tonometry is still the gold standard for measuring the intraocular pressure, its limitations in postkeratoplasty eyes, due to postoperative modified corneal morphology, have led to the search for more suitable alternatives.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!