Article Abstract

Purpose: To optimize the dose delivery to the breast lumpectomy target treated with intraoperative electron beam radiotherapy (IOERT).

Materials And Methods: Two tools have been developed in our MU calculation software NEMO X to improve the dose homogeneity and the in-vivo dosimetry effectiveness for IOERT treatments. Given the target (tumor bed) thickness measured by the surgeon, NEMO X can provide auto dose normalization to cover 95% of the target volume with 95% of the prescription dose (PD) and a "best guess" of the expected dosimeter dose (EDD) for a deep seated in-vivo dosimeter. The tools have been validated with the data of 91 patients treated with IOERT on a LIAC mobile accelerator. In-vivo dosimetry has been performed with microMOSFETs positioned on the shielding disk inserted between the tumor bed and the chest wall.

Results: On average the auto normalization showed to provide better results if compared to conventional normalization rules in terms of mean target dose (|MTD-PD|/PD ≤ 5% in 95% vs. 53% of pts) and V107 percentage (left angle bracket V107 right angle bracket =19% vs. 32%). In-vivo dosimetry MOSFET dose (MD) showed a better correlation with the EDD guessed by our tool than just by assuming that EDD=PD (|MD-EDD|/EDD ≤ 5% in 57 vs. 26% of pts).

Conclusions: NEMO X provides two useful tools for the on-line optimization of the dose delivery in IOERT. This optimization can help to reduce unnecessary large over-dosage regions and allows introducing reliable action levels for in-vivo dosimetry.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.radonc.2012.01.009DOI Listing

Publication Analysis

Top Keywords

in-vivo dosimetry
16
on-line optimization
8
intraoperative electron
8
electron beam
8
beam radiotherapy
8
dose
8
dose delivery
8
tumor bed
8
angle bracket
8
in-vivo
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!