Objective: To compare the effective radiation dose levels of cone beam computed tomography (CBCT) with those of multi-slice computed tomography (MSCT) when scanning the same maxillofacial regions.

Methods: The effective doses of 2 CBCT (NewTom 9000 and DCT Pro) and 1 MSCT (bright speed edge select 8 slice) scanners were calculated using thermoluminescent dosimeters (TLD) that were placed in a head and neck phantom, and expressed according to the International Commission on Radiation Protection (ICRP) 2007 guidelines.

Results: Effective dose values ranged from 41.8 to 249.1 µSv for CBCT. The doses of MSCT scanning for maxilla, mandible and maxilla + mandible were 506.7, 829.9 and 1066.1 µSv, respectively. Dose levels of scanning only for maxilla or mandible were significantly lower than those for maxilla + mandible.

Conclusions: When scanning the same maxillofacial regions, the dose levels for NewTom 9000 and DCT Pro CBCT images were lower than those for Bright speed edge select 8 slice MSCT images. Dose levels reduction could be obtained when smaller regions were scanned.