Purpose: Theoretical reasoning suggests that direct conversion digital x-ray detectors based upon photoconductive amorphous-selenium (a-Se) could attain very high values of the MTF (modulation transfer function) at spatial frequencies well beyond 20 cycles mm(-1). One of the fundamental factors affecting resolution loss, particularly at x-ray energies just above the K-edge of selenium (12.66 keV), is the K-fluorescence reabsorption mechanism, wherein energy can be deposited in the detector at locations laterally displaced from the initial x-ray interaction site. This paper compares measured MTF changes above and below the Se K-edge of a CCD based a-Se x-ray detector with theoretical expectations.
Methods: A prototype 25 μm sampling pitch (Nyquist frequency = 20 cycles mm(-1), 200 μm thick a-Se layer based x-ray detector, utilizing a specialized CCD readout device (200 × 400 area array), was used to make edge images with monochromatic x-rays above and below the K-edge of Se. A vacuum double crystal monochromator, exposed to polychromatic x-rays from a synchrotron, formed the monochromatic x-ray source. The monochromaticity of the x-rays was 99% or better. The presampling MTF was determined using the slanted edge method. The theory modeling the MTF performance of the detector includes the basic x-ray interaction physics in the a-Se layer as well as effects related to the operation of the CCD and charge trapping at a blocking layer present at the CCD/a-Se interface.
Results: The MTF performance of the prototype a-Se CCD was reduced from the theoretical value prescribed by the basic Se x-ray interaction physics, principally by the presence of a blocking layer. Nevertheless, the K-fluorescence reduction in the MTF was observed, approximately as predicted by theory. For the CCD prototype detector, at five cycles mm(-1), there was a 14% reduction of the MTF, from a value of 0.7 below the K-edge of Se, to 0.6 just above the K-edge.
Conclusions: The MTF of an a-Se x-ray detector has been measured using monochromatic x-rays above and below the K-edge of selenium. The MTF is poorer above the K-edge by an amount consistent with theoretical expectations.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1118/1.3673957 | DOI Listing |
J Comput Assist Tomogr
January 2025
Department of Radiology, College of Medicine, University of Florida, Gainesville, FL.
Purpose: This study evaluated beam quality and radiation dosimetry of a CT scanner equipped with a novel detector and filtration technology called PureVision Optics (PVO). PVO features miniaturized electronics, a detector cut with microblade technology, and increased filtration in order to increase x-ray detection and reduce image noise.
Methods: We assessed the performance of two similar 320-detector CT scanners: one equipped with PVO and one without.
J Comput Assist Tomogr
January 2025
Department of Radiology, College of Medicine, University of Florida, Gainesville, FL.
Purpose: The purpose of this work was to evaluate the image quality of a commercial CT scanner equipped with a novel detector and filtration technology called PureVision Optics (PVO).
Methods: CT number, noise, contrast-to-noise ratio (CNR), modulation transfer function (MTF), and noise power spectrum (NPS) were assessed using the ACR CT Accreditation phantom scanned with various acquisitions at 80 kV, 100 kV, 120 kV, and 135 kV, each with multiple CTDIvol values of 20 mGy, 40 mGy, and 65 mGy. Artifacts were evaluated in an anthropomorphic head phantom, a cadaver head, and in patient studies.
Synthetic aperture X-ray ghost imaging (SAXGI) is proposed to achieve megapixel X-ray ghost imaging together with a reduced number of measurements. As the bucket detector array is artificially generated by post-pixel-binning of the images collected with the same detector as that in the reference arm, the unique advantages of SAXGI are not verified experimentally. In this paper, we developed a systematic solution of the experimental implementation of SAXGI, with the automatic interchange of 2× and 20× optical magnification of the detector for object and reference signal acquisition respectively, together with electronic pixel-binning of the detector.
View Article and Find Full Text PDFWe demonstrate PaCMAN, a ptychography algorithm that can reconstruct high quality images with broadband illumination sources while being robust to shot, detector, and parasitic noise. We extend prior monochromatization work to improve accuracy, especially for discrete spectra, and also demonstrate how PaCMAN can be converted into Ms. PaCMAN, a multi-spectral variant that outperforms multi-spectral ePIE.
View Article and Find Full Text PDFMed Phys
January 2025
Department of Physics and Astronomy, University of California - Irvine, Irvine, California, USA.
Background: K-edge subtraction (KES) imaging is a dual-energy imaging technique that enhances contrast by subtracting images taken with x-rays that are above and below the K-edge energy of a specified contrast agent. The resulting reconstruction spatially identifies where the contrast agent accumulates, even when obscured by complex and heterogeneous distributions of human tissue. This method is most successful when x-ray sources are quasimonoenergetic and tunable, conditions that have traditionally only been met at synchrotrons.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!