Background: Cleaning of flexible endoscopes is most commonly performed using manual methods that are often performed inadequately. The aim of this study was to validate the sample collection protocol and the Rapid Use Scope Test (RUST) and then assess its usefulness in clinical use.

Methods: The benchmarks for adequate cleaning were protein <6.4 μg/cm(2), hemoglobin <2.2 μg/cm(2), and carbohydrate <1.2 μg/cm(2). Sample collection consisted of flushing 10 mL of sterile reverse osmosis water through the suction-biopsy port to the distal end. Validation of the RUST audit tool included simulated-use and in-use testing in 43 endoscopy clinics across Canada.

Results: Simulated-use testing validated that improperly cleaned endoscopes that exceeded the cleaning benchmarks would be flagged by the RUST test. The clinical-use study indicated that 96.6% of 1,489 scope channels tested were RUST negative; however, 19% and 12% of elevator guide-wire channels and endoscopic retrograde colangiopancreatography channels, respectively, exceeded the benchmarks. The survey indicated that reprocessing personnel valued a rapid audit tool for assessing compliance with manual cleaning.

Conclusion: The validated RUST test provides health care users with a rapid audit tool for manual cleaning that can be integrated into the quality program in endoscopy.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajic.2011.10.006DOI Listing

Publication Analysis

Top Keywords

rapid scope
8
scope test
8
cleaning flexible
8
development validation
4
validation rapid
4
test strips
4
strips determine
4
determine efficacy
4
efficacy manual
4
manual cleaning
4

Similar Publications

Double bowtie design for high sensitivity pediatric spectral CT.

Conf Proc Int Conf Image Form Xray Comput Tomogr

August 2024

Department of Radiology, Perelman School of Medicine, Philadelphia, PA USA.

Despite the evident benefits of spectral computed tomography (CT) in delivering qualitative imaging superior to that of conventional CT in adults, its application in pediatric diagnostic imaging is still relatively limited due to various reasons, including design limitations and radiation dose considerations. The use of specialized K-edge filters, in conjunction with other spectral technologies, has been demonstrated to improve spectral quantification accuracy. X-ray flux limitations generally pose challenges in these concepts when applied to adults.

View Article and Find Full Text PDF

Background: Migraine represents a chronic neurological disorder characterized by high prevalence, substantial disability rates, and significant economic burden. Its pathogenesis is complex, and there is currently no cure. The rapid progress in multi-omics technologies has provided new tools to uncover the intricate pathological mechanisms underlying migraine.

View Article and Find Full Text PDF

We report the rapid synthesis of primary amides by directly using commercially available ammonia borane (NH·BH), sodium hexamethyldisilazide (NaHMDS), and esters. The success of this protocol relies on NH·BH as the nitrogen source being considerably more convenient and NaHMDS being an excellent proton abstractor but not participating in the nucleophilic addition reaction. The reaction had a wide substrate scope containing bioactive molecules, and most of the substrates were efficiently amidated over 90% yields.

View Article and Find Full Text PDF

Background:  Tuberculosis (TB) remains a leading cause of mortality in low-resource settings and poses a diagnostic challenge in human immunodeficiency virus (HIV)-negative populations because of limitations in traditional diagnostic methods such as sputum smear microscopy (SSM) and sputum Xpert Ultra. There is a lack of effective, non-invasive diagnostic options for TB diagnosis in HIV-negative populations. This scoping review explores the potential of urinary lipoarabinomannan (ULAM) as a point-of-care diagnostic tool for Mycobacterium tuberculosis (MTB) in HIV-negative individuals.

View Article and Find Full Text PDF

Trace contaminants are toxic and their widespread presence in the environment potentially threatens human health. The levels of these pollutants are often difficult to determine directly using instruments owing to the complexities of environment matrices. Hence, pretreatment steps, such as sample purification and concentration, are key along with various processes that enhance the accuracy and sensitivity of the detection method.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!