Purpose: The purpose of this study was to observe and compare any loosening of screws attaching several interchangeable abutments to internally connected implants after cyclic loading.

Materials And Methods: Four different abutment groups mated with Straumann single-stage transmucosal implants (n = 7 each group) were assessed: Straumann solid abutment, Restore COC abutment, Neoplant solid abutment, and AVANA solid abutment. Each implant was fixed rigidly in a special holding jig. Abutments were tightened to 35 Ncm with a torque controller. A cyclic load of 150 N at a 30-degree angle to the long axis was applied to the implants for 1 million cycles. Prior to loading, Periotest values (PTVs) were measured. After cyclic loading, PTVs were measured and removal torque values (RTVs) of abutments were measured with a digital torque gauge.

Results: No mechanical failures were noted for the Straumann solid abutment or the Restore COC abutment. Six Neoplant abutment screws fractured (86%), and four implants fractured (57%) in the group restored with AVANA solid abutments. The RTVs of the Straumann solid abutment were significantly higher than those of the other abutments. The final mean PTV of the Straumann solid abutment was significantly lower than the final mean PTVs of the other abutments. The final mean PTV (4.76 ± 5.58) was significantly higher than the initial mean PTV (-4.29 ± 0.47).

Conclusions: Although different abutments are interchangeable with each other, they possess different chemical compositions and physical characteristics. The use of an abutment and implant manufactured by the same company is recommended to prevent the loosening of the abutment screw.

Download full-text PDF

Source

Publication Analysis

Top Keywords

solid abutment
24
straumann solid
16
abutment
12
abutments
8
interchangeable abutments
8
abutments internally
8
internally connected
8
connected implants
8
implants cyclic
8
cyclic loading
8

Similar Publications

Reusing healing abutments is common practice among clinicians; however, ensuring complete surface decontamination is crucial to avoid further complications. This study aimed to evaluate the cleaning potential of Er: YAG laser at different frequencies, as an adjunctive step prior to autoclave sterilization. Forty contaminated healing abutments were divided into four groups.

View Article and Find Full Text PDF

In essence the resin-bonded fixed partial denture, or resin-bonded bridge, is an artificial tooth that is bonded to 1 or 2 abutment teeth. If the resin-bonded bridge is directly made with resin composite intra-orally, then the connector area is the weakest link. The alternative is to produce an indirect resin-bonded bridge in which a metal or ceramic frame spans the space between the abutments.

View Article and Find Full Text PDF

Wear at the male-female interface of retentive elements in implant-supported removable prostheses is the most frequent complication in such applications. The lack of an ideal/optimal insertion path, as well as the fabrication inaccuracies, are the primary contributors to this issue. A male attachment with a common ball anchor enhanced by lateral flexibility was investigated as a solution, compared to the widely used rigid ball anchor design.

View Article and Find Full Text PDF

Objective: The purpose was to compare the biomechanical behavior of single-piece post-core restorations made from polyaryletherketone materials with fiber post-core restorations when serving as abutments for RPD using finite element analysis (FEA).

Methods: Phantom maxillary central incisor and mandibular second premolar were trimmed 1-mm coronally to cemento-enamel junction; root canals were enlarged and the teeth were scanned. Data was transferred to a solid modeling software.

View Article and Find Full Text PDF

Comparison of the reverse scan technique with an intraoral scanner and the traditional impression technique.

J Prosthet Dent

September 2024

Professor, Department of Dentistry, Charles University, 2nd Medical Faculty and Faculty Hospital Motol, Prague, Czech Republic. Electronic address:

Statement Of Problem: Intraoral scanners have many advantages but have limited applicability for extensive tooth-supported or implant-supported prostheses because of merging errors.

Purpose: The purpose of this study was to compare the reverse scan technique (RST) with an intraoral scanner using the traditional impression technique both in vitro and in vivo.

Material And Methods: A participant was scanned 10 times with an intraoral scanner.

View Article and Find Full Text PDF