Aims: Caused by ageing of the population, better survival from ischaemic heart disease, and improved treatment of chronic heart disease, the incidence of heart failure has increased enormously. Worldwide, left ventricular assist devices (LVADs) are increasingly being used as a bridge or alternative to heart transplantation. In this study, we investigated whether there is difference in functional and haemodynamic recovery after implantation of pulsatile and continuous-flow pumps.
Methods And Results: We compared laboratory and echocardiographic data and exercise performance in patients with end-stage heart failure, before and 3 months after implantation of pulsatile and continuous-flow LVADs. A significant improvement in all laboratory parameters after implantation of both types of LVADs was seen, as well as a significant decrease in heart rate and LV dimensions, indicating better haemodynamics and cardiac recompensation. This improvement was better for the pulsatile device, probably due to higher plasma levels and higher LV dimensions before implantation. Exercise capacity strongly improved: 3 months after implantation of pulsatile and continuous-flow LVADs, peak VO(2) was 20.2 ± 4.8 vs. 18.3 ± 4.8 mL/kg/min (P = 0.09) (53 ± 12 vs. 49 ± 11% of predicted for age and gender) (P = 0.28).
Conclusion: Pulsatile and continuous-flow LVADs result in extensive haemodynamic recovery and exercise performance compatible with daily life activities. Exercise performance with continuous-flow LVADs is equal to that with pulsatile devices. This, in combination with improved survival of the newer devices, allows its use as an alternative to heart transplantation in selected patients.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1093/eurjhf/hfr181 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!