Nanoparticle penetration through filter media and leakage through face seal interface of N95 filtering facepiece respirators.

Ann Occup Hyg

Policy and Standard Development Branch, National Institute for Occupational Safety and Health, National Personal Protective Technology Laboratory, 626 Cochrans Mill Road, PO Box 18070, Pittsburgh, PA 15236, USA.

Published: July 2012

AI Article Synopsis

  • The National Institute for Occupational Safety and Health (NIOSH) recommends using particulate respirators to protect against nanoparticles, particularly those smaller than 100 nm.
  • A study has shown that the total inward leakage (TIL) of nanoparticles is influenced by both filter penetration and leakage through the respirator's face seal, with a specific focus on particles around 45 nm being the most penetrating size.
  • Research involving various N95 respirator models revealed that filter penetration increased with smaller particles (like 45 nm) and higher breathing volumes, demonstrating how both filter efficiency and artificial leaks at the face seal impact overall respiratory protection.

Article Abstract

National Institute for Occupational Safety and Health recommends the use of particulate respirators for protection against nanoparticles (<100 nm size). Protection afforded by a filtering facepiece particulate respirator is a function of the filter efficiency and the leakage through the face-to-facepiece seal. The combination of particle penetration through filter media and particle leakage through face seal and any component interfaces is considered as total inward leakage (TIL). Although the mechanisms and extent of nanoparticle penetration through filter media have been well documented, information concerning nanoparticle leakage through face seal is lacking. A previous study in our laboratory measured filter penetration and TIL for specific size particles. The results showed higher filter penetration and TIL for 50 nm size particles, i.e. the most penetrating particle size (MPPS) than for 8 and 400 nm size particles. To better understand the significance of particle penetration through filter media and through face seal leakage, this study was expanded to measure filter penetration at sealed condition and TIL with artificially introduced leaks for 20-800 nm particles at 8-40 l minute volumes for four N95 models of filtering facepiece respirators (FFRs) using a breathing manikin. Results showed that the MPPS was ~45 nm for all four respirator models. Filter penetration for 45 nm size particles was significantly (P < 0.05) higher than the values for 400 nm size particles. A consistent increase in filter penetrations for 45 and 400 nm size particles was obtained with increasing breathing minute volumes. Artificial leakage of test aerosols (mode size ~75 nm) through increasing size holes near the sealing area of FFRs showed higher TIL values for 45 nm size particles at different minute volumes, indicating that the induced leakage allows the test aerosols, regardless of particle size, inside the FFR, while filter penetration determines the TIL for different size particles. TIL values obtained for 45 nm size particles were significantly (P < 0.05) higher than the values obtained for 400 nm size particles for all four models. Models with relatively small filter penetration values showed lower TIL values than the models with higher filter penetrations at smaller leak sizes indicating the dependence of TIL values on filter penetration. When the electrostatic charge was removed, the FFRs showed a shift in the MPPS to ~150 nm with the same test aerosols (mode size ~75 nm) at different hole sizes and breathing minute volumes, confirming the interaction between filter penetration and face seal leakage processes. The shift in the MPPS from 45 to 150 nm for the charge removed filters indicates that mechanical filters may perform better against nanoparticles than electrostatic filters rated for the same filter efficiency. The results suggest that among the different size particles that enter inside the N95 respirators, relatively high concentration of the MPPS particles in the breathing zone of respirators can be expected in workplaces with high concentration of nanoparticles. Overall, the data obtained in the study suggest that good fitting respirators with lower filter penetration values would provide better protection against nanoparticles.

Download full-text PDF

Source
http://dx.doi.org/10.1093/annhyg/mer122DOI Listing

Publication Analysis

Top Keywords

nanoparticle penetration
4
penetration filter
4
filter media
4
media leakage
4
leakage face
4
face seal
4
seal interface
4
interface n95
4
n95 filtering
4
filtering facepiece
4

Similar Publications

Developing Topics.

Alzheimers Dement

December 2024

IMDEA Nanociencia, Madrid, Spain.

Background: About half of the patients suffering from Alzheimer's disease (AD) display sleeping disorders. Disruptions in the central circadian clock (CC), located in the brain, accelerate AD pathogenesis, making the CC a promising target. In preclinical trials, this strategy have shown efficacy but clinical results are inconsistent.

View Article and Find Full Text PDF

Drug Development.

Alzheimers Dement

December 2024

MEPSGEN, Seoul, Korea, Republic of (South).

Background: Impaired Aβ clearance plays a key role in the common, late-onset AD. Anti-Aβ immunotherapies are controversial, in part because of high rates of serious side effects including edema, microhemorrhages, and siderosis, highlighting the importance of the development of alternative Aβ clearance strategy. Here, we introduce a bioinspired nanoparticle named MG-PE3 crossing the human blood-brain barrier (BBB) and clearing Aβ with no adverse effect.

View Article and Find Full Text PDF

Background: Cognitive decline associated with Alzheimer's disease (AD) correlates with hyperphosphorylated tau (pTau) propagating between neurons along networks connected by synapses. It has been hypothesized this transcellular transmission occurs partially by extracellular vesicles (EVs). Both genetic and pharmacological inhibition of nSMase2 has been found to inhibit EV biogenesis and pTau propagation.

View Article and Find Full Text PDF

Introduction Endodontic re-infections primarily occur due to the ingress of bacteria and their toxins through an incomplete seal following obturation. A variety of sealers have been developed to achieve effective integration with the different obturation materials and dentinal tubules. To choose the right endodontic sealer and application for each clinical instance, one must be aware of the attributes of the various sealers commonly used in clinical practice.

View Article and Find Full Text PDF

Efficient Cytosolic Delivery of Single-Chain Polymeric Artificial Enzymes for Intracellular Catalysis and Chemo-Dynamic Therapy.

J Am Chem Soc

January 2025

The State Key Laboratory of Molecular Engineering of Polymers and Department of Macromolecular Science, Fudan University, Shanghai 200438, P. R China.

Designing artificial enzymes for in vivo catalysis presents a great challenge due to biomacromolecule contamination, poor biodistribution, and insufficient substrate interaction. Herein, we developed single-chain polymeric nanoparticles with Cu/N-heterocyclic carbene active sites (SCNP-Cu) to function as peroxidase mimics for in vivo catalysis and chemo-dynamic therapy (CDT). Compared with the enzyme mimics based on unfolded linear polymer scaffold and multichain cross-linked scaffold, SCNP-Cu exhibits improved tumor accumulation and CDT efficiency both in vitro and in vivo.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!