AI Article Synopsis

  • A new technique for inserting distal locking screws in intramedullary nails reduces both operating time and radiation exposure compared to traditional methods.
  • The guided freehand approach uses computer-assisted projections to help surgeons accurately drill into bone, relying only on standard C-arm imaging equipment.
  • Testing on human cadaver legs showed a significant reduction in the number of images taken and radiation time, highlighting its efficiency and safety benefits for patients and surgical teams.

Article Abstract

Background: The common practice for insertion of distal locking screws of intramedullary nails is a freehand technique under fluoroscopic control. The process is technically demanding, time-consuming and afflicted to considerable radiation exposure of the patient and the surgical personnel. A new concept is introduced utilizing information from within conventional radiographic images to help accurately guide the surgeon to place the interlocking bolt into the interlocking hole. The newly developed technique was compared to conventional freehand in an operating room (OR) like setting on human cadaveric lower legs in terms of operating time and radiation exposure.

Methods: The proposed concept (guided freehand), generally based on the freehand gold standard, additionally guides the surgeon by means of visible landmarks projected into the C-arm image. A computer program plans the correct drilling trajectory by processing the lens-shaped hole projections of the interlocking holes from a single image. Holes can be drilled by visually aligning the drill to the planned trajectory. Besides a conventional C-arm, no additional tracking or navigation equipment is required.Ten fresh frozen human below-knee specimens were instrumented with an Expert Tibial Nail (Synthes GmbH, Switzerland). The implants were distally locked by performing the newly proposed technique as well as the conventional freehand technique on each specimen. An orthopedic resident surgeon inserted four distal screws per procedure. Operating time, number of images and radiation time were recorded and statistically compared between interlocking techniques using non-parametric tests.

Results: A 58% reduction in number of taken images per screw was found for the guided freehand technique (7.4 ± 3.4) (mean ± SD) compared to the freehand technique (17.6 ± 10.3) (p < 0.001). Total radiation time (all 4 screws) was 55% lower for the guided freehand technique compared to conventional freehand (p = 0.001). Operating time per screw (from first shot to screw tightened) was on average 22% reduced by guided freehand (p = 0.018).

Conclusions: In an experimental setting, the newly developed guided freehand technique for distal interlocking has proven to markedly reduce radiation exposure when compared to the conventional freehand technique. The method utilizes established clinical workflows and does not require cost intensive add-on devices or extensive training. The underlying principle carries potential to assist implant positioning in numerous other applications within orthopedics and trauma from screw insertions to placement of plates, nails or prostheses.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3305668PMC
http://dx.doi.org/10.1186/1471-2474-13-8DOI Listing

Publication Analysis

Top Keywords

freehand technique
28
guided freehand
20
conventional freehand
16
freehand
12
technique compared
12
compared conventional
12
operating time
12
technique
9
distal interlocking
8
radiation exposure
8

Similar Publications

Objective: Accurate rotational reduction following tibial shaft fracture fixation is absent in up to 36% of cases yet may be critical for lower extremity biomechanics. The objective of this cadaveric study was to compare the results of freehand methods of reduction with software-assisted reduction.

Methods: Four fellowship-trained orthopaedic trauma surgeons attempted rotational correction in a cadaveric model with fluoroscopic assistance (without radiographic visualization of the fracture site) using (1) their method of choice (MoC) and (2) software assistance (SA).

View Article and Find Full Text PDF

Background: Recent advances within the last decade have allowed robotics to become commonplace in the operating room. In the field of neurosurgery, robotics assist surgeons in pedicle screw placement and vertebral fusion procedures. The purpose of this review is to look at currently used spinal robots available on the market and compare their overall accuracy, cost, radiation exposure, general adverse events, and hospital readmission rates.

View Article and Find Full Text PDF

Robot-assisted medial patellofemoral ligament reconstruction in the treatment of recurrent patellar dislocation can improve tunnel accuracy but yields similar outcome compared with traditional technique.

Arthroscopy

January 2025

Department of Orthopaedics, Lanzhou University Second Hospital, Lanzhou City, Gansu Province, China; Orthopaedics Clinical Medical Research Center of Gansu Province, Lanzhou University Second Hospital, Lanzhou City, Gansu Province, China; Intelligent Orthopaedic Industry Technology Center of Gansu Province, Lanzhou University Second Hospital, Lanzhou City, Gansu Province, China. Electronic address:

Purpose: To review patients with recurrent patellar dislocation surgically treated with robot-assisted medial patellofemoral ligament (MPFL) reconstruction compared with patients who underwent surgery using the traditional freehand technique.

Methods: A retrospective cohort study was performed to identify patients who underwent MPFL reconstruction from January 2020 to December 2023 in our hospital. The inclusion criteria were: patients aged from 15 to 50 years; patellar dislocation occurred two or more times; a Merchant view or computed tomography (CT) scan indicating patellofemoral joint malalignment, external patellar inclination, or lateral patellar dislocation; underwent MPFL reconstruction via robot-assisted or traditional freehand technique; complete medical records and imaging data before and after surgery; a minimum of 1 year of postoperative follow-up.

View Article and Find Full Text PDF

Purpose: The aim of this study was to evaluate the feasibility of using patient-specific implants (PSI) for complex shaft corrective osteotomies in multiplanar deformities of long bones in the lower extremities. Additionally, it aimed to investigate the added value of these implants by quantifying surgical accuracy on postoperative CT, comparing their outcomes to two commonly used techniques: 3D virtual visualizations and 3D-printed surgical guides.

Methods: Six tibial and femoral shaft corrective osteotomies were planned and performed on three Thiel embalmed human specimen.

View Article and Find Full Text PDF

Purpose: To investigate the feasibility of a robotic system with artificial intelligence-based lesion detection and path planning for CT-guided biopsy, compared to the conventional freehand technique.

Materials And Methods: Eight nodules within an abdominal phantom, incorporating the simulated vertebrae and ribs, were designated as targets. A robotic system was used for lesion detection, trajectory generation, and needle-holder positioning.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!