Dose uniformity analysis among ten 16-slice same-model CT scanners.

J Comput Assist Tomogr

Diagnostic X-Ray Quality Assurance Section, Department of Medical Physics, Memorial Sloan-Kettering Cancer Center (MSKCC), New York, NY 10065, USA.

Published: April 2012

Purpose: With the introduction of multislice scanners, computed tomographic (CT) dose optimization has become important. The patient-absorbed dose may differ among the scanners although they are the same type and model. To investigate the dose output variation of the CT scanners, we designed the study to analyze dose outputs of 10 same-model CT scanners using 3 clinical protocols.

Materials And Methods: Ten GE Lightspeed (GE Healthcare, Waukesha, Wis) 16-slice scanners located at main campus and various satellite locations of our institution have been included in this study. All dose measurements were performed using poly (methyl methacrylate) (PMMA) head (diameter, 16 cm) and body (diameter, 32 cm) phantoms manufactured by Radcal (RadCal Corp, Monrovia, Calif) using a 9095 multipurpose analyzer with 10 × 9-3CT ion chamber both from the same manufacturer. Ion chamber is inserted into the peripheral and central axis locations and volume CT dose index (CTDIvol) is calculated as weighted average of doses at those locations. Three clinical protocol settings for adult head, high-resolution chest, and adult abdomen are used for dose measurements.

Results: We have observed up to 9.4% CTDIvol variation for the adult head protocol in which the largest variation occurred among the protocols. However, head protocol uses higher milliampere second values than the other 2 protocols. Most of the measured values were less than the system-stored CTDIvol values. It is important to note that reduction in dose output from tubes as they age is expected in addition to the intrinsic radiation output fluctuations of the same scanner.

Conclusion: Although the same model CT scanners were used in this study, it is possible to see CTDIvol variation in standard patient scanning protocols of head, chest, and abdomen. The compound effect of the dose variation may be larger with higher milliampere and multiphase and multilocation CT scans.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RCT.0b013e31823ed149DOI Listing

Publication Analysis

Top Keywords

dose
10
same-model scanners
8
dose output
8
ion chamber
8
adult head
8
ctdivol variation
8
head protocol
8
protocols head
8
higher milliampere
8
scanners
7

Similar Publications

Objective: Within the scope of this research, the long-term effects of experimental blunt head trauma on immature rats and MK-801 administered acutely after trauma on the brain tissue will be examined. In addition, the impact of trauma and MK-801 on Nestin and CD133, which are essential stem cells, will be evaluated by immunohistochemical and ELISA methods.

Methods: In this study, the contusion trauma model was used.

View Article and Find Full Text PDF

Target cyclooxygenase 2 (COX-2) and 5-lipoxygenase (5-LOX) inhibitors; 5-([2,5-Dihydroxybenzyl]amino)salicylamides (Compounds 1-11) were examined for potential anticancer activity, with a trial to assess the underlying possible mechanisms. Compounds were assessed at a single dose against 60 cancer cell lines panel and those with the highest activity were tested in the five-dose assay. COMPARE analysis was conducted to explore potential mechanisms underlying their biological activity.

View Article and Find Full Text PDF

Impact of Norepinephrine Use on Free Flap Survival in Breast Reconstructive Microsurgery.

Microsurgery

January 2025

Service de Chirurgie Plastique et Reconstructrice, Hôpital européen Georges-Pompidou, Paris, France.

Objective: The optimal method for maintaining intraoperative blood pressure during microsurgical procedures remains controversial. While intravenous fluid administration is essential, overfilling can lead to complications. Vasopressor agents are used cautiously due to their vasoconstrictive effects, which could potentially lead to flap failure.

View Article and Find Full Text PDF

The determinants of varicella-zoster virus (VZV)-associated central nervous system (CNS) infection have not been fully elucidated. This study aimed to investigate the incidence and risk factors, including immunosuppression, for different manifestations of VZV-associated CNS infection. Patient registers were used to include adults diagnosed with VZV-associated CNS infections between 2010 and 2019 in Sweden.

View Article and Find Full Text PDF

Fluoride Exposure Modulates Skeletal Development and Mineralization in Zebrafish Larvae.

Environ Toxicol

January 2025

Molecular Toxicology Laboratory, Department of Biotechnology, Bharathiar University, Coimbatore, Tamil Nadu, India.

The presence of high levels of fluoride (F) in groundwater is a major issue worldwide. Although F is essential for healthy teeth and bones, excessive exposure can cause fluorosis or F toxicity. This condition primarily affects the hard tissues due to their high F retention capacity.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!