Objective: To determine differences in coronary artery calcium (CAC) measurement performed with the use of 2 generations of multidetector computed tomography (CT) scanners of the same manufacturer.

Methods: Agatston Score (AS) and calcium mass (CM) were measured with a 4-row scanner (AS4 and CM4) and a 64-row scanner (AS64 and CM64) using a cardiac phantom with calcium inserts.

Results: The results of the AS measurements (mean ± SD) varied significantly between the equipment: 880.6 ± 30.1 (AS4) vs 586.5 ± 24.0 (AS64; P < 0.0001). The AS interscanner variability was 31.6% for the phantom and from 25.5% to 110.1% for particular inserts. Mean ± SD CM values were different as well: 192.8 ± 5.0 mg (CM4) vs 152.4 ± 2.6 mg (CM64; P < 0.0001). Determination of CM with 64-row CT was more accurate than that with an older scanner; the mean relative error was -9.1% and 15.0%, respectively (P < 0.0001). The CM interscanner variability was 23.3% for the phantom and from 19.0% to 122.8% for particular inserts. The interexamination variability ranged from 1.7% (CM64) to 5.6% (AS4).

Conclusions: Coronary artery calcium scoring with the 64-row CT scanner is more accurate than with the 4-row device The difference between the results of AS and CM measurements carried out with both scanners is statistically significant.

Download full-text PDF

Source
http://dx.doi.org/10.1097/RCT.0b013e31823d796cDOI Listing

Publication Analysis

Top Keywords

calcium scoring
8
computed tomography
8
coronary artery
8
artery calcium
8
64-row scanner
8
00001 interscanner
8
interscanner variability
8
comparison calcium
4
scoring 4-multidetector
4
4-multidetector computed
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!