Electrical Impedance Tomography (EIT) can resolve dynamic physiological information deep within human subjects [1], but its sensitivity is challenged in the case of imaging the head [2]. Here, we report a new system called fEITER that has been designed and built to enable functional imaging of the human brain using EIT via scalp-mounted electrodes, integrated with stimulation of evoked responses. Using Field-Programmable Gate Array (FPGA) technology, it provides excellent flexibility in terms of current-pattern excitation and signal processing. The instrument operates at 100 frames/second (fps) with noise of 1 μV on the rms voltage measurements. Clinical trials have been authorized by the UK MHRA and example data from human subjects are presented.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1109/IEMBS.2011.6091777 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!