Flexible PET/ITO electrode array for implantable biomedical applications.

Annu Int Conf IEEE Eng Med Biol Soc

Integrated Circuits and Systems Laboratory, Electrical & Computer Eng Dept, KN Toosi University of Technology, Tehran, Iran.

Published: June 2012

Flexible PET/ITO (PolyEthylene Terephthalate/ Indium Tin Oxide) implantable electrode array for spinal cord stimulation and retina prosthesis have been developed. The electrode array is fabricated on a thin PET/ITO substrate and encapsulated with insulating material, SU-8. The PET substrate made electrodes flexible so that they could shape to contoured tissues. A layer of gold on the stimulation sites served to reduce the electrode/tissue interface impedance. Prototypes of 1 × 8 and 3 × 8 electrode arrays are fabricated for monophasic and biphasic stimulation of spinal cord respectively. The exposed electrode dimensions are 3 mm² for monophasic and 6 mm² for biphasic stimulation with 100 μm of interconnection paths. The prototype of 4 × 4 electrode arrays were also fabricated with the same process for retinal prosthesis with exposed electrode diameter of 125 μm. To verify the functionality of the subdural electrodes, the electrochemical impedance spectroscopy was measured. The electrode/tissue impedance was 500 Ω at 1KHz for 3 mm² area.

Download full-text PDF

Source
http://dx.doi.org/10.1109/IEMBS.2011.6090794DOI Listing

Publication Analysis

Top Keywords

electrode array
12
flexible pet/ito
8
spinal cord
8
electrode arrays
8
arrays fabricated
8
biphasic stimulation
8
exposed electrode
8
electrode
7
pet/ito electrode
4
array implantable
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!