Background: Urinary dipsticks are the most frequent method used for screening of ketones in animals, but this method has many drawbacks. In human medicine, portable meters that measure ketones in whole blood have largely replaced urinary dipsticks.
Objective: The aim of this prospective study was to validate a portable whole-blood ketone meter for use in cats.
Methods: Sixty-two cats (11 clinically healthy, 51 with diabetes mellitus) were included in the study. The concentration of β-hydroxybuyrate (β-HB) was measured in venous and capillary blood with a hand-held ketone meter (Precision Xceed; assay range 0-8 mmol/L) and compared with a spectrophotometric method. Precision, accuracy, and the effects of hematocrit and anticoagulants were evaluated.
Results: Between-run precision using low- and high-concentration control solutions was 8.1% and 2.6%, respectively; within-run coefficient of variation determined using 12 feline blood samples was 2.8%. In the 62 cats, β-HB concentrations measured with the portable ketone meter ranged from 0-7.4 mmol/L (median 0.9 mmol/L). When β-HB concentrations measured by the portable meter were < 4.0 mmol/L there was good agreement with the reference method, but concentrations > 4.0 mmol/L were lower than those obtained by the reference method in 20 of 24 cats (83%). There was good correlation between capillary and venous measurements. Results were not affected by hematocrits from 0.17 to 0.50 L/L, but EDTA was not a suitable anticoagulant.
Conclusion: Measurement of β-HB concentration in peripheral or capillary blood by an easy-to-use portable ketone meter was suitable for detecting ketonemia in cats. Underestimation of β-HB concentration was observed at higher values, but results were sufficiently high to aid in diagnosing diabetic ketoacidosis.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1939-165X.2011.00389.x | DOI Listing |
BMC Oral Health
November 2024
Prosthodontic Department, Faculty of Dentistry, Tanta University, Tanta, Egypt.
Background: The accuracy of digital implant transfer is currently under investigation in relation to the effect of saliva, scan body material, and exposed length.
Methods: Six completely edentulous casts with four implant fixtures were fabricated. The four implant fixtures in each cast were placed below the crest of the ridge of the casts by 1.
J Oral Biol Craniofac Res
October 2024
Department of Prosthodontics and Crown & Bridge, SRM Dental College, Bharathi Salai, Ramapuram, Chennai, TamilNadu, India.
Purpose: This study evaluated the influence of different agents such as blood, artificial saliva, and normal saline on preload force of dental implants with bio-high-performance poly-ether-ether-ketone (Bio-HPP) abutments to determine its effect on screw loosening.
Methods: Forty (N = 40) Grade 5 titanium dental implant analog (GM Implant Analog; Neodent, Straumann) with Bio-HPP poly ether-ether ketone (PEEK) abutment and titanium screw was used in the study. The samples were embedded in acrylic split mold.
Diabet Med
September 2024
Norwich Medical School, University of East Anglia, Norwich, Norfolk, UK.
Aim: To determine whether it was feasible, safe and acceptable for ambulance clinicians to use capillary blood ketone meters for 'high-risk' diabetic ketoacidosis (DKA) recognition and fluid initiation, to inform the need for a full-powered, multi-centre trial.
Methods: Adopting a stepped-wedge controlled design, participants with hyperglycaemia (capillary blood glucose >11.0 mmol/L) or diabetes and unwell were recruited.
J Chromatogr A
August 2024
Department of Instrumentation and Analytical Chemistry, CAS Key Laboratory of Separation Science for Analytical Chemistry, Dalian Institute of Chemical Physics, Chinese Academy of Sciences (CAS), Dalian 116023, PR China; Technology Innovation Center of Food Safety Technique of Inspection for State Market Regulation (Rapid Screening and Traceability for Edible Agricultural Product Safety), PR China. Electronic address:
In this work, a microchip gas chromatography (GC) column assembly utilizing a three-dimensional (3D) printed micro oven and a flexible stainless steel capillary column was developed. The assembly's performance and separation capabilities were characterized. The key components include a 3D printed aluminum plate (7.
View Article and Find Full Text PDFACS Sens
February 2024
Department of Nanoengineering, University of California San Diego, La Jolla, California 92093, United States.
Ketone bodies (KBs), especially β-hydroxybutyrate (BHB), have gained tremendous attention as potential biomarkers as their presence in bodily fluids is closely associated with health and wellness. While a variety of blood fingerstick test strips are available for self-testing of BHB, there are major needs for wearable devices capable of continuously tracking changing BHB concentrations. To address these needs, we present here the first demonstration of a wearable microneedle-based continuous ketone monitoring (CKM) in human interstitial fluid (ISF) and illustrate its ability to closely follow the intake of ketone drinks.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!