Objective: To evaluate the effect that a safety wire (SW) had on insertion force and direct ureteral trauma using ureteral access sheaths (UAS) in an ex vivo porcine model.

Material And Methods: UASs were advance into 20 fresh ex vivo porcine kidney-ureters, either with or without SWs. The average and max force of insertion were recorded. Retrograde saline injection was used to identify the presence of lacerations.

Results: For SW vs no SW, the mean insertion force (1.79 vs 0.67, P = .0003, respectively) and max insertion force (2.29 kg vs 1.00, P = .0007) was greater in the SW group. There were 9 lacerations, 6 of which were partial and 3 complete. Of the partial injuries, 2 were in the no SW group and 4 were in the SW group. Of the complete lacerations, 2 occurred in the no SW group and 1 in the SW group. However, there was no significant difference in the total number of lacerations (P = 1.00), the number of partial lacerations (P = .628), or the number of complete lacerations (P = 1.00) between the 2 groups.

Conclusions: The use of an SW significantly increases the force required to insert a UAS; however, it does not appear to affect the force or the degree of ureteral injury.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.urology.2011.11.002DOI Listing

Publication Analysis

Top Keywords

insertion force
16
vivo porcine
12
ureteral injury
8
ureteral access
8
access sheaths
8
group group
8
complete lacerations
8
lacerations 100
8
force
7
ureteral
5

Similar Publications

Background & Objective: Currently, there are many implants in clinical use, making it hard to choose the right one for the patient. The success rate of an implant depends on its diameter, length, and direction of insertion in bone. In implant dentistry, Finite Element Analysis (FEA) simulates intraoral conditions in vitro and analyzes the effects of implant material, diameter, size, and other components related to oral structure on the implant and peri-implant tissues.

View Article and Find Full Text PDF

Orthopedic screws are widely used to achieve bone reduction, compression, and construct stability. However, the relationship between insertion torque, interfragmentary compression, and fixation strength, especially when comparing standard screws with NiTiNOL/sustained dynamic compression (SDC), has not been thoroughly investigated. This study measured insertion torque, interfragmentary compression, and fixation strength for two types of headed orthopedic devices-standard and SDC-using solid foam bone replicates and cadaver validation.

View Article and Find Full Text PDF

High-affinity VNARs targeting human hemoglobin: Screening, stability and binding analysis.

Int J Biol Macromol

January 2025

College of Ocean Food and Biological Engineering, Jimei University, Xiamen 361021, China. Electronic address:

Hemoglobin, composed of α- and β-chains, is essential for oxygen transport and is key in diagnosing and treating gastrointestinal and blood disorders. It also aids in detecting blood contamination and estimating transfusion volumes. Immunological methods, based on antigen-antibody interactions, are distinguished by their high sensitivity and accuracy.

View Article and Find Full Text PDF

Eccentric Reaming is Superior to Augmented Components in B2 Glenoids: a Biomechanical study.

J Shoulder Elbow Surg

January 2025

Division of Orthopaedic Surgery, The Ottawa Hospital, University of Ottawa, Ottawa, ON, Canada. Electronic address:

Introduction: Primary glenohumeral arthritis is typically associated with glenoid retroversion and posterior bone loss. Glenoid component fixation remains a weak link in the survivorship of anatomical total shoulder arthroplasty, particularly in the B2 glenoid. The aim of this study was to compare biomechanical properties of two glenoid preparation techniques in a B2 glenoid bone loss model.

View Article and Find Full Text PDF

What Is the Optimal Geometry of Dissolving Microneedle Arrays? A Literature Review.

Pharmaceutics

January 2025

Department of Pharmaceutical Technology and Biopharmacy, University of Groningen, 9713 AV Groningen, The Netherlands.

The application of dissolving microneedle arrays (DMNAs) is an emerging trend in drug and vaccine delivery as an alternative for hypodermic needles or other less convenient drug administration methods. The major benefits include, amongst others, that no trained healthcare personnel is required and that the recipient experiences hardly any pain during administration. However, for a successful drug or vaccine delivery from the DMNA, the microneedles should be inserted intact into the skin.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!