Purpose: The efficient use of computed tomography (CT) and magnetic resonance imaging (MRI) equipment necessitates establishing adequate quality-control (QC) procedures. In particular, the accuracy of slice thickness (ST) requires scan exploration of phantoms containing test objects (plane, cone or spiral). To simplify such procedures, a novel phantom and a computerised LabView-based procedure have been devised, enabling determination of full width at half maximum (FWHM) in real time.
Materials And Methods: The phantom consists of a polymethyl methacrylate (PMMA) box, diagonally crossed by a PMMA septum dividing the box into two sections. The phantom images were acquired and processed using the LabView-based procedure.
Results: The LabView (LV) results were compared with those obtained by processing the same phantom images with commercial software, and the Fisher exact test (F test) was conducted on the resulting data sets to validate the proposed methodology.
Conclusions: In all cases, there was no statistically significant variation between the two different procedures and the LV procedure, which can therefore be proposed as a valuable alternative to other commonly used procedures and be reliably used on any CT and MRI scanner.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s11547-011-0775-5 | DOI Listing |
J Biomed Opt
January 2025
CIFICEN (UNCPBA - CICPBA - CONICET), Tandil, Argentina.
Significance: In the last years, time-resolved near-infrared spectroscopy (TD-NIRS) has gained increasing interest as a tool for studying tissue spectroscopy with commercial devices. Although it provides much more information than its continuous wave counterpart, accurate models interpreting the measured raw data in real time are still lacking.
Aim: We introduce an analytical model that can be integrated and used in TD-NIRS data processing software and toolkits in real time.
Conf Proc Int Conf Image Form Xray Comput Tomogr
August 2024
Department of Radiology, Perelman School of Medicine, Philadelphia, PA, USA.
Respiratory motion phantoms can be used for evaluation of CT imaging technologies such as motion artifact reduction algorithms and deformable image registration. However, current respiratory motion phantoms do not exhibit detailed lung tissue structures and thus do not provide a realistic testing environment. This paper presents PixelPrint, a method for 3D-printing deformable lung phantoms featuring highly realistic internal structures, suitable for a broad range of CT evaluations, optimizations, and research.
View Article and Find Full Text PDFJ Vasc Surg Cases Innov Tech
April 2025
Department of Radiology and Nuclear Medicine, University Teaching and Research Hospital Lucerne, Luzern, Switzerland.
Objective: The aim of this single-center case series is to demonstrate that an ultra-low dose (ULD) can be routinely achieved in the hybrid operating room in standard endovascular aortic repair (EVAR) for infrarenal abdominal aortic aneurysm by adjusting the manufacturer's predefined imaging parameters, hardware configurations and user protocols (including benchmarking).
Methods: The hybrid operating room manufacturer predefined EVAR software setup of the dose exposure control software (OPTIQ, Siemens Healthineers, Forchheim, Germany) at our university medical center was screened for possible improvements regarding radiation dose application. Tests on a water-equivalent as well as polymethyl methacrylate phantom model to assess the impact of technical settings were performed, including comparison of settings for exposure control software, different magnification, collimation configurations and detector distance.
Med Phys
January 2025
Department of Medical Biophysics, Schulich School of Medicine and Dentistry, Western University, London, Ontario, Canada.
Background: With increasing evidence supporting three-dimensional (3D) automated breast (AB) ultrasound (US) for supplemental screening of breast cancer in increased-risk populations, including those with dense breasts and in limited-resource settings, there is an interest in developing more robust, cost-effective, and high-resolution 3DUS imaging techniques. Compared with specialized ABUS systems, our previously developed point-of-care 3D ABUS system addresses these needs and is compatible with any conventional US transducer, which offers a cost-effective solution and improved availability in clinical practice. While conventional US transducers have high in-plane resolution (axial and lateral), their out-of-plane resolution is constrained by the poor intrinsic elevational US resolution.
View Article and Find Full Text PDFRadiat Oncol
January 2025
Department of Radiotherapy, Changzhou Cancer Hospital, Honghe Road, Xinbei Area, Changzhou, 213032, China.
Purpose: Conventional radiotherapy (CRT) has limited local control and poses a high risk of severe toxicity in large lung tumors. This study aimed to develop an integrated treatment plan that combines CRT with lattice boost radiotherapy (LRT) and monitors its dosimetric characteristics.
Methods: This study employed cone-beam computed tomography from 115 lung cancer patients to develop a U-Net + + deep learning model for generating synthetic CT (sCT).
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!