Purpose: A novel patient-specific intensity modulated radiation therapy (IMRT) QA system, 3DVH software and mapcheck 2, purports to be able to use diode array-measured beam doses and the patient's DICOM RT plan, structure set, and dose files to predict the delivered 3D dose distribution in the patient for comparison to the treatment planning system (TPS) calculated doses. In this study, the composite dose to an ion chamber and film in phantom predicted by the 3DVH and mapcheck 2 system is compared to the actual measured chamber and film doses. If validated in this context, then 3DVH can be used to perform an equivalent dose analysis as that obtained with film dosimetry and ion chamber-based composite IMRT QA. This is important for those losing their ability to perform film dosimetry for true composite IMRT QA and provides a measure of confidence in the accuracy of 3DVH 3D dose calculations which may replace phantom-based IMRT QA.

Methods: The dosimetric results from 15 consecutive patient-specific IMRT QA tests performed by composite field irradiation of ion chamber and EDR2 film in a solid water phantom were compared to the predicted doses for those virtual detectors based on the calculated 3D dose by the 3DVH software using mapcheck 2 measured doses of each beam within each plan. For each of the 15 cases, immediately after performing the ion chamber plus film measurements, the mapcheck 2 was used to measure the dose for each beam of the plan. The dose to the volume of the virtual ion chamber and the dose distribution in the plane of the virtual film calculated by the 3DVH software was extracted. The ratio of the measured to 3DVH or eclipse-predicted ion chamber doses was calculated. The same plane in the phantom measured using film and calculated with eclipse was exported from 3DVH and the 2D gamma metric was used to compare the relationship between the film doses and the eclipse or 3DVH predicted planar doses. Also, the 3D gamma value was calculated in the 3DVH software which compares the eclipse dose to the 3DVH predicted dose distribution. For the 2D and 3D gamma metrics, 2% dose and 2 mm distance to agreement (DTA) were used. In addition, a simple dose difference was performed using either a 2% or 3% dose difference tolerance.

Results: The mean ratio ± standard deviation of the measured vs 3DVH or vs eclipse-predicted dose to the ion chamber was 1.013 ± 0.015 and 1.003 ± 0.012, respectively. For 3DVH vs eclipse, the mean percentage of pixels failing the 3D gamma metric was 1.2% ± 1.4% while the failure rate for the 2D gamma metric was 1.1% ± 0.9%. When either 3DVH or eclipse was compared to EDR2 film, the gamma failure rate was 2.3% ± 2.0% and 1.6% ± 1.7%, respectively. Mean dose difference failures were 9%-27% ± 5%-15% for 2 or 3% dose difference tolerances, depending on the combination of systems tested. No statistically significant differences were found for any of the planar dosimetric comparisons.

Conclusions: 3DVH + mapcheck 2 predicts the same absolute dose, the percent of pixels failing the gamma metric, and the percent of pixels failing 2% or 3% dose difference tolerance tests as one would have obtained had one made measurements in solid water phantom using an ion chamber and coronal film instead of a diode array. This is also a necessary although not sufficient condition for validation of the accuracy of 3DVH predictions of the 3D dose using beam-by-beam measurements.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.3666771DOI Listing

Publication Analysis

Top Keywords

ion chamber
32
dose
21
3dvh software
20
dose difference
20
3dvh
16
chamber film
16
gamma metric
16
accuracy 3dvh
12
film
12
composite imrt
12

Similar Publications

Despite the importance of ocular surface in human physiology and diseases, little is known about ion channel expression, properties and regulation in ocular epithelial cells. Furthermore, human primary epithelial cells have rarely been studied in favor of rat, mouse and especially rabbit animal models. Here, we developed primary human Meibomian gland (hMGEC) and conjunctival (hConEC) epithelial cells.

View Article and Find Full Text PDF

TiCT/Au NPs/PPy ternary heterostructure-based intra-capacitive self-powered sensor for DEHP detection.

J Hazard Mater

January 2025

Department of Analytical Chemistry, College of Chemistry, Jilin University, Changchun 130012, China. Electronic address:

Phthalate esters, particularly di(2-ethylhexyl) phthalate (DEHP), are widely used plasticizers found in various consumer products, posing significant environmental and health risks due to their endocrine-disrupting effects. In this study, a novel enzyme-free intra-capacitive biofuel cell self-powered sensor (ICBFC-SPS) was developed. The ICBFC-SPS integrated a ternary heterostructure-based capacitive anode and a cathode with a sensing interface into a single-chamber electrolytic cell.

View Article and Find Full Text PDF

Autism spectrum disorder (ASD) is linked to ion channel dysfunction, including chloride voltage-gated channel-4 (CLCN4). We generated Clcn4 knockout (KO) mice by deleting exon 5 of chromosome 7 in the C57BL/6 mice. Clcn4 KO exhibited reduced social interaction and increased repetitive behaviors assessed using three-chamber and marble burying tests.

View Article and Find Full Text PDF

The durability of reinforced concrete is associated with several factors that can trigger the corrosion of reinforcement bars. Among these factors, the most significant are chloride-ion attack and carbonation. This study evaluated, through accelerated testing, self-compacting concretes (SCCs) with reduced cement content in binary, ternary, and quaternary mixtures using high-early-strength Portland cement, fly ash (FA), metakaolin (MK), and hydrated lime (HL).

View Article and Find Full Text PDF

The integrity of esophageal epithelial cells in patients with gastroesophageal reflux disease (GERD) or GERD-like symptoms is the first mechanism of protection to decrease the sensitivity to gastric reflux and heartburn symptoms. We investigated the protective effects of Poliprotect (PPRO), a CE-marked medical device, on esophageal epithelial integrity using in vitro and ex vivo models. In vitro, the protective effects of PPRO were tested on Caco-2 cells.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!