Purpose: To individually benchmark the incident electron parameters in a Monte Carlo model of an Elekta linear accelerator operating at 6 and 15 MV. The main objective is to establish a simplified but still precise benchmarking procedure that allows accurate dose calculations of advanced treatment techniques.
Methods: The EGSnrc Monte Carlo user codes BEAMnrc and DOSXYZnrc are used for photon beam simulations and dose calculations, respectively. A 5 × 5 cm(2) field is used to determine both the incident electron energy and the electron radial intensity. First, the electron energy is adjusted to match the calculated depth dose to the measured one. Second, the electron radial intensity is adjusted to make the calculated dose profile in the penumbrae region match the penumbrae measured by GafChromic EBT film. Finally, the mean angular spread of the incident electron beam is determined by matching calculated and measured cross-field profiles of large fields. The beam parameters are verified for various field sizes and shapes.
Results: The penumbrae measurements revealed a non-circular electron radial intensity distribution for the 6 MV beam, while a circular electron radial intensity distribution could best describe the 15 MV beam. These electron radial intensity distributions, given as the standard deviation of a Gaussian distribution, were found to be 0.25 mm (in-plane) and 1.0 mm (cross-plane) for the 6 MV beam and 0.5 mm (both in-plane and cross-plane) for the 15 MV beam. Introducing a small mean angular spread of the incident electron beam has a considerable impact on the lateral dose profiles of large fields. The mean angular spread was found to be 0.7° and 0.5° for the 6 and 15 MV beams, respectively.
Conclusions: The incident electron beam parameters in a Monte Carlo model of a linear accelerator could be precisely and independently determined by the benchmarking procedure proposed. As the dose distribution in the penumbra region is insensitive to moderate changes in electron energy and angular spread, accurate penumbra measurements is feasible for benchmarking the electron radial intensity distribution. This parameter is particularly important for accurate dosimetry of mlc-shaped fields and small fields.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1118/1.3668315 | DOI Listing |
J Chem Phys
January 2025
Theoretical and Computational Physics Section, Raja Ramanna Centre for Advanced Technology, Indore 452013, India.
The orbital-free density functional theory (OF-DFT) based method is a convenient tool to carry out electronic structure calculations scaling almost linearly with the number of electrons. However, the main impediment in the application of this method is the unavailability of the accurate form for the non-interacting kinetic energy functional in terms of electron density. The Pauli kinetic energy functional is the unknown part of the kinetic energy functional, and the corresponding Pauli potential appears in the governing Euler equation.
View Article and Find Full Text PDFAppl Radiat Isot
January 2025
School of Artificial Intelligence, Wenzhou Polytechnic, Wenzhou, 325035, China. Electronic address:
For the purpose of assessing image quality and calculating patient X-ray dosage in radiology, computed tomography (CT), fluoroscopy, mammography, and other fields, it is necessary to have prior knowledge of the X-ray energy spectrum. The main components of an X-ray tube are an electron filament, also known as the cathode, and an anode, which is often made of tungsten or rubidium and angled at a certain angle. At the point where the electrons generated by the cathode and the anode make contact, a spectrum of X-rays with energies spanning from zero to the maximum energy value of the released electrons is created.
View Article and Find Full Text PDFJ Mol Model
January 2025
Department of Chemistry, Birla Institute of Technology and Science Pilani - K. K. Birla Goa Campus, Zuarinagar, 403726, Goa, India.
Context: There has been growing interest in amino acid ionic liquids because of their low-cost synthesis and superior biodegradability and biocompatibility compared to traditional ionic liquids. In this study, we have investigated the structure and dynamics of three ionic liquids consisting of N-butyl N-methyl piperidinium [Pip] cation with amino acid (lysine [Lys], histidine [His], and arginine [Arg]) anions. The radial distribution functions, the spatial distribution functions, and the coordination numbers have been used to analyze the structure in the bulk phase.
View Article and Find Full Text PDFInorg Chem
January 2025
School of Materials Science and Engineering, China University of Petroleum, Qingdao 266580, PR China.
Mismatched electron and proton transport rates impede the manifestation of effective performance of the electrocatalytic oxygen evolution reaction (OER), thereby limiting its industrial applications. Inspired by the natural protein cluster in PS-II, different organic-inorganic hybrid electrocatalysts were synthesized via a hydrothermal method. -Toluidine (PT), benzoic acid (BA), and -aminobenzoic acid (PABA) were successfully intercalated into NiFe-LDH.
View Article and Find Full Text PDFThe present study demonstrates the significance of the C/N ratio and double helical ribbon (DHR) impeller in the anaerobic co-digestion (AnCo-D) of sugar refining process (SRP) effluent and molasses-based distillery spent wash (DSW) for improved biogas production. Both SRP & DSW were mixed in different percentages to achieve an optimum C/N ratio. Further biomethane potential analysis of mixed feeds with different C/N ratios was performed.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!