Topiramate (Topamax®) is an antiepileptic medication used as adjunctive and monotherapy in patients with epilepsy and for migraine prophylaxis. A GC-MS assay was developed that was capable of detecting topiramate plasma concentrations following a single rectal or oral dose administration. Topiramate plasma samples were prepared by solid-phase extraction and were quantified by GC-MS analysis. The topiramate standard curves were split from 0.1 to 4 µg/mL and from 4 to 40 µg/mL in order to give a more accurate determination of the topiramate concentration. The accuracy of the standards ranged from 94.6 to 107.3% and the precision (%CV) ranged from 1.0 to 5.3% for both curves at all concentrations. The %CV for quality controls was <7.6%. The assay is both accurate and precise and will be used to complete future pharmacokinetic studies.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC3776594PMC
http://dx.doi.org/10.1002/bmc.1749DOI Listing

Publication Analysis

Top Keywords

topiramate plasma
12
analysis topiramate
8
topiramate
6
sensitive capillary
4
capillary gc-ms
4
gc-ms method
4
method analysis
4
plasma single-dose
4
single-dose studies
4
studies topiramate
4