The purpose of this pilot study was to test the feasibility of a wireless 5-sensor, motion detection system (5S-MDS) with hospitalized older adults. Interventions to prevent hospital-based falls in older adults are important to reduce morbidity, mortality, and health care costs. Wearable motion sensors, which track and wirelessly transmit body movements, may identify human movement patterns that immediately precede falls, thus allowing early prevention. Descriptive feasibility study in which 5 hospitalized older adults were recruited to wear the 5S-MDS for 4 hours. Measurement included assessment of participant acceptance, skin integrity, and sensor accuracy. All 5 participants (mean age, 90.2 years) agreed that sensors were acceptable and skin integrity was maintained. The sensor data accurately reflected the patient movements. The 5S-MDS was feasible for 4 hours' use with hospitalized older adults. It has potential as an early warning system for falls.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.gerinurse.2011.11.011DOI Listing